Intelect® Mobile Stim
PATiENT PrEPArATiON
37
ELECTrOTHErAPy PATiENT PrEPArATiON (continued)
Insert the lead with the Red (+) electrode
connector into one Dura-Stick Electrode.
Insert the lead with the Black (-) electrode
connector into the other electrode.
Make certain the Lead Wires are
seated completely into the electrodes.
NOTE:
Use of conductive medium or
sponges is not required or recommended.
Dura-Stick Electrodes are manufactured to
ensure the optimum conductivity during
therapy when properly applied.
Red (+)
LeAd WIRe
bLACk (-)
LeAd WIRe
DUra-sTiCK Electrode instructions
Connecting lead Wires
Remove the Dura-Stick Electrodes from
the protective backing and apply to the
treatment area as prescribed. Ensure the
entire electrode surface is in contact with
patient skin by pressing into place.
securing Electrodes
LeAd WIRe SeATed