ARM PREHEAT TIMER
Plug the unit back into an approved power
source. If the “Preheat Timer” is used, press
the “PREHEAT TIMER” button once, the
“PREHEAT TIMER ON” indicator will illuminate,.
in order for the Preheat Timer to automatically
come on at the default time
the next morning.
NOTE:
The Fluidotherapy Preheat Timer does not
operate on Saturday or Sunday. However,
for Monday morning preheat, set the unit as
described above on Friday evening.
REFILL WITH CELLEX® MEDIUM
Refill the unit with Cellex® Dry Heat Medium to
approximately one (1) inch above the bottom of
the arm treatment ports of the unit.
NOTE:
Use of other than Cellex Medium may cause
premature failure of the fluidotherapy unit(s).
22
Fluidotherapy®- Dry Heat Therapy
REQUIRED MAINTENANCE
DAILY MAINTENANCE
CLEAN INLET FILTERS
At the end of each work day, unplug the
unit and clean the Inlet Filters on the unit.
Carefully remove the filter retainer and
wash the filter and screen with a mild
antibacterial soap and water. Thoroughly
dry the filter and screen before placing
back on the unit.
NOTE:
Should your unit have the earlier style inlet
filters, gently clean the filter using a soft
bristle brush. Be careful not to puncture or
damage the filter.
Should the filter become damaged, torn or
punctured, call your dealer for replacement
of the filter before resuming operation.
•
Before any Maintenance is performed or attempted, unplug the unit from
the power source to prevent the possibility of electrical shock.
WARNING
INSPECT SLEEVE CONDITION
Inspect the port sleeves for tears, rips and weak seams. Replace all sleeves
that show signs of tears, rips, weak or loose seams or excessive wear.
Keeping the sleeves in excellent condition prevents excessive spillage
of the Cellex medium and prevents the medium from entering the heat
chamber of the unit.