CAUTION!
When you press the START key, the device automatically enters
the programmed movement range. The device stops when it
reaches the preprogrammed range of motion. To start treatment,
program the device if necessary (see
Programming the Device
section, page 11), then restart.
ADJUSTING THE OPTIFLEX
®
S
Numbers in parenthesis are for assistance with parts identification. References can
be found in the chapter entitled
Parts Description,
on pages 22 and 23 of this manual.
1. First connect the plug to a safety socket and switch the Main Switch (7) on.
2. Insert an original OptiCard
™
(2) into the pendant (1).
3. Activate the programming mode by pressing the
AB/ADDUCTION
and
STOP
buttons at the same time, or hold down the STOP button for five seconds.
4. Move the arm support into a position comfortable for the patient (abduction,
adduction, internal/external rotation or ante/retroversion).
5. Carry out setting from steps number one through five (see numbering on device).
Ante/Retroversion (Horizontal Abduction/Adduction) (Figure A)
Ante/retroversion is adjusted manually. The aim of the adjustment is to find the most
comfortable position for the patient.
Carry out this adjustment using the Adjustment Knob (11). Tighten the Adjustment
Knob securely after adjustment.
Height Adjustment (Figures B and C )
1. Secure the arm support at the outer tube for upper arm strength (16).
2. Loosen wing screw (13). Align the axis of Motor A (15) on the patient’s shoulder
joint pivot (see Figure C1). Make sure the patient is in a straight, upright position.
3. The pivot of Motor A (15) must correspond to the pivot of the patient’s shoulder
joint (see Figure C2).
4. Tighten the wing screw (13) securely.
A
B
C
OPTIFLEX S USER MANUAL
C1
C2
CAUTION!
Make sure that the rotating axis of the device (Motor B) and the
rotating axes of the shoulder joint correspond for
abduction/adduction and rotation (see Figure C).
NOTE: The plugs for the mobility element can only be connected in
one direction (see figure D). Also, make sure that the mobility
element’s connector is always locked on using the safety hinge.