OPTIFLEX S USER MANUAL
12
Programming Steps, Continued
5. Press the STOP button to store all previous values.
6 Pressing the parameter buttons while the device is stopped shows the currently
stored values.
7. Press the START button. The settings will be automatically checked by the
device (settings are shown in the display). After initiating the programmed
Range of Motion, the device will stop at a point midway between the beginning
and the ending range of motion to allow the patient comfortable access.
8. Press the START button again to start the treatment.
N
NO
OT
TE
E::
The desired motion dimensions must be at least 5° different.
Otherwise, the “DIFF” error code will appear on the LED Display.
Information on the Treatment Values
Maximum settings of the motion ranges:
Adduction
30°
Abduction
175°
Elevation (flexion)
30° - 175°
(Only to be used in isolation without rotation.
Set Motor B in 90° external rotation and switch off.)
Internal Rotation
90°
External Rotation
90°
Manual
Anteversion (horizontal adduction)
120°
Retroversion (horizontal abduction)
0°
CAUTION!
Recommended sequence for programming:
1. Adduction
2. Internal Rotation
3. External Rotation
4. Abduction
Unless otherwise prescribed by the supervising doctor.
CAUTION!
Differences might occur between the angle specifications on
the device and the angles setting of the extremity. Painless,
relaxed movement of the extremity is of paramount
importance.
OPTIFLEX S USER MANUAL
CAUTION!
When using a programmed patient chip card, press S
STTA
AR
RTT
after the card has been entered without the patient. The
motion device automatically moves into the motion range
and stops. Press S
STTA
AR
RTT again to begin treatment.
CAUTION!
If the shoulder immobilization option is used, abduction may
not be programmed above 80°.