Typical Applications
23
Press
【
Sweep
】
key and press
[Power]
in the soft key menu displayed in the right side of screen.
Press
[Power Level]
in submenu ,press
【
+/-
】
→
【
10
】
→
【
/off
】
in input box displayed in the upper
right side of screen and set up current power as -10dBm.
Maximum power 3656 Series Vector Network Analyzer can test is +10dBm. It is required to
lower output power of the analyzer because the gain of amplifier tested is around 12dB. It can
be guaranteed that the maximum power will not be exceeded after going through amplifier. If
gain of the amplifier is considerably large, then it shall be considered to test by connecting
attenuator at the end of output end of the amplifier.
3.2.2
Calibration
To improve the accuracy of measurement results, you may calibrate the analyzer before measuring the
DUT to eliminate system errors. It is required to perform dual-port SOLT calibration for the measurement
of an amplifier. Calibration kit 20202(3.5mm) is selected suitable for the DUT. Refer to calibration chapter
3.1.2 in
Filter Test
above mentioned for calibration process.
3.2.3
Measurement of DUT
Cut off cable between Port 1 and Port 2 after completing calibration. Perform measurement by
connecting amplifier in the middle, as shown in the following picture.
Figure 3-20 Amplifier Connection Diagram
Notice