5.2 Detector - Functions L
F
, L
S
and L
I
For electrical calibration, use the ADM0C130 adapter.
Greatest peak-to-peak voltage applicable with the
ADM0C130 adapter
26
Vpp
Maximum error in the maximum response to a tone burst
Function
Duration of the tone burst (ms)
Maximum error (dB)
L
F
200
± 1.0
L
S
500
±
1.0
L
I
20
5
2
± 1.5
± 2.0
± 2.0
Maximum error for signals with crest factor
≤
3:
± 0.5 dB
Maximum error for signals with crest factor
≤
5:
± 1.0 dB
Maximum error for signals with crest factor
≤
10:
± 1.5 dB
Maximum overshot:
L
F
:
L
S
:
1.1 dB
1.6 dB
Maximum error of level linearity (31.5 to 12500 Hz):
± 0.7 dB
Maximum error of differential level linearity (31.5 to 12500
Hz):
± 0.2 dB
Peak detector - L
I
function
Decay rate:
2.9 dB/s
±
0.5 dB/s
Onset time constant:
< 3.5 ms
5.3 Peak detector - L
peak
function
Onset time constant:
< 75
µ
s
5.4 Integrator - L
T
, L
t
and L
E
functions
Linearity range:
110 dB
Pulse range:
65 dB
Response time to a steady input signal:
2 s