30
may continue to consult the functions measured prior to the temporary interruption. To
resume measurement, press
.
When you press
measurement will stop. The state of measurement indicator will
change from to .
2.3.5 Consulting measured data
While the SC310 is in stop mode ( ) you may consult all the functions measured so far.
To consult them, follow the same procedure as described in the section 2.3.3 on displaying
data while measurement is in progress.
When measurement has stopped ( ) you may display the value of the total percentiles
(belonging to the total measurement time t). The statistical sound-meter mode screen
displays the following information:
•
Value of the total percentiles L
1
, L
5
, L
10
, L
50
, L
90
, L
95
and L
99
.
•
Total measurement time t.
2.4 Measuring in spectrum analyser mode 1/1
The spectrum analyser mode 1/1 is ideal for conducting real-time measurement of sound-
pressure levels, both frequential values for octave bands centred on frequencies 31.5, 63,
125, 250, 500, 1000, 2000, 4000, 8000 and 16000 Hz (without frequency weighting), and
Overall values with all frequency weightings (A, C and Z). The SC310 also measures the
peak sound pressure level for the consecutive integration time T with all frequency
weightings (A, C and Z). The SC310 measures all these functions simultaneously, in real
time, together with the 'short' levels (125 ms) and seven percentiles for each of the octave
bands as well as the overall percentiles with frequency weighting A. The SC310 features a
room background noise evaluation screen based on the NC (Noise Criterion) curves.
Major applications of this mode include measurement of soundproofing, frequential analysis
of industrial, environmental and workplace noise, analysis of noise generated by air-
conditioning systems and room background noise.
2.4.1 Prior adjustments: integration time
Before beginning to measure in spectrum analyser mode, the following parameters should be
set:
•
The integration time (T) used in the evaluation of spectral and overall levels. This
integration is carried out in consecutive T time intervals.