9.12 Monitoring test progress
Fig. 18
– Exam screen while test is in progress
The following commands are available:
1. Pupil images from alignment cameras allow the
operator to check if subject’s eyes are open
2. Pupil size:
if smaller than 3 mm a “small pupil” warning is displayed. Pupil size may oscillate
during the test due to accommodation
3. Allows to adjust the height of the chin rest
4. Live retinal image with stimuli allows the operator to monitor progress. Fovea should stay at
the center of the gray cross. Seen stimuli are shown as filled dots. The stimulus being
projected is shown as an empty circle
5. Tracking status (circular colored frame, see par. 9.11)
6. Elapsed test time
7. Completed stimuli over total being projected: test proceeds slow at the beginning and turns
progressively faster
8. Allows to stop the test at any time
COMPASS requires a minimum pupil of 3.0 mm: a test with average pupil size smaller
than 3.0 mm may be unreliable.
1
2
3
4
5
6
7
8