![CEM Smart Trac II Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/cem/smart-trac-ii/smart-trac-ii_quick-start-manual_2580966028.webp)
Data In Review
Edit References:
1. 12.20
2. 10.56
3. 5.30
Edit Standard – 12.20%
%Fat
S/M
Mass
------------
----- ----------
*
1. 12.24 1539.6340 2.2315 g
*
2. 12.15 1534.0634 2.3152 g
*
3. 12.19 1537.0634 2.5315 g
Edit Standard – 10.56%
%Fat
S/M
Mass
------------
----- ----------
*
1. 10.56 1414.2478 2.4311 g
*
2. 10.48 1409.0634 2.3315 g
*
3. 10.61 1417.2236 2.5314 g
Edit Standard – 5.30%
%Fat
S/M
Mass
------------
----- ----------
*
1. 5.26 1036.0111 2.1222 g
*
2. 5.35 1040.1713 2.2231 g
*
3. 5.30 1038.0225 2.1315 g
Once appropriate points are selected and there are at least 3 points for each reference
within specification, exit to Main Menu (saves method). Then,
EXPORT METHODS
to a
USB stick (refer to REFERENCE PROCEDURES). Please,
METHODS
to your
CEM Representative for back up.