![CEM Smart Trac II Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/cem/smart-trac-ii/smart-trac-ii_quick-start-manual_2580966026.webp)
Create a Fat NMR Method on the SMART Trac II
From the
Main Menu
Select
#2
. Edit/Create Method
Select
#1
New Method
Name Method: (spell out name of method); Select
(“Ready”)
Edit Method
#1
.
Moisture/Fat NMR
(“toggle”)
Caution:
Do Not Choose
Moisture/Fat
this selection is for the FAS-9001
#2. Constant Weight
(
Default
)
#3.
Turbo
ON (
Default
)
Moisture/Fat NMR Type
#1.
Standard
Enter Parameters for Moisture
:
Power:
100%
Delta Weight:
0.2mg
Delta
Time:
10sec
Max
Time:
10min
(
Default
)
Bias:
+0
(
Default
)
Min
Result:
0%
(
Default
)
Max
Result:
100%
(
Default
)
(“Next Page”)
Max
Temp:
100
o
C
Based on AOAC oven method
Min Wt Range:
2.0g
(2.0 is
Default
)
Max Wt Range:
2.5g
Wt
Compensation:
ON
(“Next Page”)
Fat Range <80%, >80% or <1%
(<80
Default
, <1% Condensed Skim)
Fat Bias:
0%
80% Butter, Mayonnaise
Warm up delay:
0
Used with Fat Only Option, powders
or
butter.
●
Run Time
8 sec.
(Default)
(“Next Page”)
Содержание Smart Trac II
Страница 1: ...Corporation Quick Start Guide 600269 Rev 1 ...
Страница 8: ...System Setup ...
Страница 9: ...SMART Trac II Setup Guide ...
Страница 12: ...3 Right click SMARTNMR and select Tools System Setup ...
Страница 13: ...4 Select on System Setup ...
Страница 16: ...SHIM Setup prompt for Oil Standard Cal lock prompt for Normalization Standard ...
Страница 17: ...Frequency Optimization Prompt for Oil Standard Pulse Width Optimization Prompt for Oil Standard ...
Страница 23: ...Method Importation ...
Страница 25: ...Method Creation ...