5.
Once the System Setup dialog window is open, select Auto Run.
“Auto Run” will automatically run the following procedures, stopping
between each procedure to prompt for user input:
• Magnet Temperature Offset
• SHIM Setup
• Cal-lock
• Frequency Optimization
• Pulse Width Optimization
Correct standard to use for each procedure:
SHIM Setup
Oil Standard
Cal-lock
Normalization Standard
Frequency Optimization
Oil Standard
Pulse Width Optimization
Oil Standard