True RMS Autoranging Digital Multimeter
2-2.Per IEC1010 Overvoltage Installation Category
Overvoltage Category I
Equipment of
Overvoltage Category I
is equipment for connection to circuits in which measures
are taken to limit the transient overvoltages to an appropriate low level.
Note:
Examples include protected electronic circuits.
Overvoltage Category II
Equipment of
Overvoltage Category II
is energy-consuming equipment to be supplied from the
fixed installation.
Note:
Examples include household, office, and laboratory appliances.
Overvoltage Category III
Equipment of
Overvoltage Category III
is equipment in fixed installations.
Note:
Examples include switches in the fixed installation and some equipment for industrial use
with permanent connection to the fixed installation.
Overvoltage Category IV
Equipment of
Overvoltage Category IV
is for use at the origin of the installation.
Note:
Examples include electricity meters and primary over-current protection equipment.
2-3.Safety Instructions
• This meter has been designed for safe use, but must be operated with caution.
• The rules listed below must be carefully followed for safe operation.
2-3-1.
N
EVER apply voltage or current to the meter that exceeds the specified maximum:
Input Protection Limits
Function
V DC
V AC
mA AC/DC
A AC/DC
Frequency, Resistance, CAP,
Diode Test, Continuity
Surge Protection: 8kV peak per IEC 61010
Maximum Input
1000VDC RMS
1000VAC RMS
800mA 1000V fast acting fuse
10A 1000V fast acting fuse (30 seconds max every 15 minutes)
250VDC/AC RMS