2
Features
Memory for 32,000 readings
Draw graphics curve by the data through the PC software
Selectable sampling interval: 2s, 5s, 10s, 30s, 1m, 5m, 10m, 30m, 1hr, 2hr, 3hr, 6hr, 12hr, 24hr
Operating status indicated by red, yellow and green LED light
Data download to PC software by USB interface
Too low/too high alarm value setup
Two modes (manual and automatic) to start recording data
Low battery alarm
Sensor: fan sensor
Operating temperature:
32 to 122
°
F (0 to 50
°
C)
Operating humidity: <80% RH
Operating height:
2000 meters (7000ft) maximum
Weight: 49g
Size: 145x 35 x 30mm
Air Velocity
Range
Resolution
Accuracy
M/S(meters per sec)
1.1 - 20.00 m/s
0.01 m/s
± (3% + 0.20 m/s)
KPH (kilometers/hour)
0.8- 72.0 km/h
0.1 km/h
± (3% + 1.0 km/hr)
FPM (feet per minute)
80
– 3937 ft/min
1 ft/min
± (3% + 40 ft/m)
MPH (miles per hour)
0.9
– 44.8 mph
0.1 mph
± (3% + 0.4 MPH)
KNT (nautical MPH)
0.8 to 38.8 knots
0.1 knots
± (3% + 0.4 knots)