EXTERNAL AIR COOLING

3. 

On the lower rear side of the Discover install the pressure regulator to the Discover using the 

3.1. 

¼” polyethylene tubing (Figure 1). 

Connect the other side of the pressure regulator to a source of compressed air and set the 

3.2. 

pressure regulator to 20 PSI.

NITROGEN SUPPLY, VACUUM, AND VENT

4. 

Connect the tubing labeled “Nitrogen” extending from the rear of the valve panel (Figure 1) to 

4.1. 

the pressure regulator of the nitrogen source.

Remove the fittings, including the yellow ferrule and nut, from the tubing attached to the 

4.2. 

vacuum pump.  Use the union fitting to connect the 1/8 in (3 mm) tube labeled “Vacuum” 

extending from the rear of the valve panel to the tubing extending from the vacuum pump.

Place the open end of the tubing labeled “Vent” extending from the rear of the valve panel into 

4.3. 

a fume hood or waste trap.

CALIBRATE FIBER-OPTIC PROBE

5. 

Connect the power cord to the Discover.  Locate the power switch (on the left side of the unit) 

5.1. 

and turn the power ON.

Press EDIT.

5.2. 

The screen should say “TEMPERATURE”.  If yes, press ENTER.  Otherwise, use the right 

5.3. 

arrow key to scroll to “TEMPERATURE” and press ENTER.

Use the right arrow key to scroll to “Select Alternate”.

5.4. 

Ensure that “DEVICE = FIBER-OPTIC”.  If yes, press ENTER.  If no, use the right arrow key to 

5.5. 

change to “DEVICE = FIBER-OPTIC” and press ENTER.

Press EDIT.

5.6. 

The screen should say “TEMPERATURE”.  If yes, press ENTER.  Otherwise, use the right 

5.7. 

arrow key to scroll to “TEMPERATURE” and press ENTER.

Highlight “Enter Calibration” and press ENTER.

5.8. 

Press EDIT, then enter the 7 digit GF# located on the white band at the base of the fiber-optic 

5.9. 

probe.

Press ENTER.  The calibration is now saved in the software.

5.10. 

Press HOME to return to the Home Screen.

5.11. 

CALIBRATE PRESSURE SENSOR

6. 

Press EDIT.

6.1. 

Use the right arrow key to scroll to “PRESSURE” and press ENTER.

6.2. 

Select “Enter Calibration” and press ENTER.

6.3. 

Press EDIT.  The last digit of SLOPE should be flashing.

6.4. 

Use the keypad to enter the Slope value printed on the white band on the collar of the 9-pin 

6.5. 

connector, then press ENTER.

The last digit of INTERCEPT should be flashing.  Use the keypad to enter the Intercept value 

6.6. 

printed on the white band, then press ENTER.  The calibration is now saved in the software.

Press HOME to return to the Home Screen.

6.7. 

INSTRUMENT OPERATION

Choose a method.

1. 

To create a new method:

1.1. 

Press the Open Folder key.

1.1.1. 

Using the left arrow, select “New Method”.

1.1.2. 

Press the right arrow key until “Mode = Standard”.  Press ENTER.

1.1.3. 

Set Power, then press ENTER.  (150 W is recommended.)

1.1.4. 

Set Maximum Temperature, then press ENTER. (150° C is recommended.)

1.1.5. 

Set Pressure, then press ENTER. (100 psi is the MAXIMUM recommended pressure.)

1.1.6. 

Содержание Discover Protein Hydrolysis

Страница 1: ...rature control 2 1 box Connect the other end of the cable to COM2 on the read of the Discover Figure 1 Connect the blue fiber optic temperature probe to the temperature control box taking care not 2 2 to bend the probe Figure 1 Instrument Connection Ports COM2 indicated in red Pressure Port indicated in yellow 600313 Rev 5 10 6 2008 NOTE The Discover Protein Hydrolysis is only compatable with Disc...

Страница 2: ... the right arrow key to 5 5 change to DEVICE FIBER OPTIC and press ENTER Press EDIT 5 6 The screen should say TEMPERATURE If yes press ENTER Otherwise use the right 5 7 arrow key to scroll to TEMPERATURE and press ENTER Highlight Enter Calibration and press ENTER 5 8 Press EDIT then enter the 7 digit GF located on the white band at the base of the fiber optic 5 9 probe Press ENTER The calibration ...

Страница 3: ...t until the top of the 2 1 thermowell is even with the top of the nut Figure 2 center picture Pass the closed end of the thermowell through the center port of the 2 port vessel cap until 2 2 the nut meets the threaded port Tighten the nut finger tight onto the threaded port to create a secure seal Figure 2 right picture Do not use a wrench as this will damage the threads Add the protein solution t...

Страница 4: ...l from the bath Thread the tubing 2 9 from the To Vessel port on the valve panel assembly through the top of the attenuator port opening Secure the end of the tubing onto the open port of the vessel using an open ferrule nut Figure 5 Place the vessel assembly into the Discover microwave cavity Secure the attenuator into 2 10 position Figure 5 Insert the fiber optic probe though the attenuator port...

Страница 5: ...d then to Run Press the PLAY key to run your selected hydrolysis program 4 At the conclusion of the hydrolysis and cooling cycle remove the temperature probe from the vessel 5 assembly Remove the attenuator then remove the vessel from the Discover cavity The vessel assembly will 6 still be connected to the valve panel Turn the valve handle to Vent to release the remaining nitrogen pressure Disconn...

Страница 6: ...4530 Fiber Optic Temperature Control 541175 Discover Protein Hydrolysis 200 240V 50 60Hz 925482 Includes the Following Bioscience Discover 200 240V 50 60Hz 908010 Discover Protein Hydrolysis Option 925590 Vacuum Pump 200 240V 50 60Hz 544540 Fiber Optic Temperature Control 541175 ACCESSORIES PRODUCT NAME PART Complete Valve Panel Assembly 908275 Reaction Vessel Top with 2 Ports 148460 Reaction Vess...

Отзывы: