RECOMMENDED PARAMETERS FOR PROTEIN HYDROLYSIS

Power =  

150 W

Power Mode =  

STANDARD

Temperature =  

150° C

Temperature Control = 

Fiber-optic

Pressure= 

100 psi

Run Time = 

2:30

Hold Time = 

15:00

PARTS AND ACCESSORIES

SYSTEM

PRODUCT NAME

PART #

Discover Protein Hydrolysis

    (100-120V/50-60Hz)

925480

 

Includes the Following:

 

  Bioscience Discover (100-120V/50-60Hz)

908005

 

  Discover Protein Hydrolysis Option

925590

 

  Vacuum Pump (100-120V/50-60Hz)

544530

 

  Fiber-Optic Temperature Control

541175

Discover Protein Hydrolysis

    (200-240V/50-60Hz)

925482

 

Includes the Following:

 

  Bioscience Discover (200-240V/50-60Hz)

908010

 

  Discover Protein Hydrolysis Option

925590

 

  Vacuum Pump (200-240V/50-60Hz)

544540

 

  Fiber-Optic Temperature Control

541175

ACCESSORIES

PRODUCT NAME

PART #

Complete Valve Panel Assembly

908275

Reaction Vessel Top with 2 Ports

148460

Reaction Vessel Bottom

324015

Microvial Tray Insert

324020-P

Condensation Shield

324030

Fiber-Optic Temperature Probe

314325

Thermowell Assembly (Qty 2)

430676

Discover Protein Hydrolysis Operations Manual

600313

Wrench Set

SP-1367

Содержание Discover Protein Hydrolysis

Страница 1: ...rature control 2 1 box Connect the other end of the cable to COM2 on the read of the Discover Figure 1 Connect the blue fiber optic temperature probe to the temperature control box taking care not 2 2 to bend the probe Figure 1 Instrument Connection Ports COM2 indicated in red Pressure Port indicated in yellow 600313 Rev 5 10 6 2008 NOTE The Discover Protein Hydrolysis is only compatable with Disc...

Страница 2: ... the right arrow key to 5 5 change to DEVICE FIBER OPTIC and press ENTER Press EDIT 5 6 The screen should say TEMPERATURE If yes press ENTER Otherwise use the right 5 7 arrow key to scroll to TEMPERATURE and press ENTER Highlight Enter Calibration and press ENTER 5 8 Press EDIT then enter the 7 digit GF located on the white band at the base of the fiber optic 5 9 probe Press ENTER The calibration ...

Страница 3: ...t until the top of the 2 1 thermowell is even with the top of the nut Figure 2 center picture Pass the closed end of the thermowell through the center port of the 2 port vessel cap until 2 2 the nut meets the threaded port Tighten the nut finger tight onto the threaded port to create a secure seal Figure 2 right picture Do not use a wrench as this will damage the threads Add the protein solution t...

Страница 4: ...l from the bath Thread the tubing 2 9 from the To Vessel port on the valve panel assembly through the top of the attenuator port opening Secure the end of the tubing onto the open port of the vessel using an open ferrule nut Figure 5 Place the vessel assembly into the Discover microwave cavity Secure the attenuator into 2 10 position Figure 5 Insert the fiber optic probe though the attenuator port...

Страница 5: ...d then to Run Press the PLAY key to run your selected hydrolysis program 4 At the conclusion of the hydrolysis and cooling cycle remove the temperature probe from the vessel 5 assembly Remove the attenuator then remove the vessel from the Discover cavity The vessel assembly will 6 still be connected to the valve panel Turn the valve handle to Vent to release the remaining nitrogen pressure Disconn...

Страница 6: ...4530 Fiber Optic Temperature Control 541175 Discover Protein Hydrolysis 200 240V 50 60Hz 925482 Includes the Following Bioscience Discover 200 240V 50 60Hz 908010 Discover Protein Hydrolysis Option 925590 Vacuum Pump 200 240V 50 60Hz 544540 Fiber Optic Temperature Control 541175 ACCESSORIES PRODUCT NAME PART Complete Valve Panel Assembly 908275 Reaction Vessel Top with 2 Ports 148460 Reaction Vess...

Отзывы: