Series VC-50 / SVC-50 / MVC-50 / LVC-50
49-MCVC50EN03
USER´S MANUAL
11
10. CALIBRATION
Switch off the indicator.
Switch them on again and during the countdown from 9 to 0, press
.
The display will show “CAL”. While the “CAL” is being shown, press
,
and
in the foregoing order in
order to enter calibration. The display will show “noLoAd”.
Remove any weight from the platform. Press
.
The display will show “LoAd”. Place the calibration weight on the scales. Press
.
The display will show the last calibration weight used. If this weight is correct, continue by pressing
. If it is
not correct, press the arrow keys to change the calibration weight value. When it is correct, press
.
If the calibration is correct the display will return to normal mode. If an error message appears, try to calibrate
the weight again since there may have been something that prevented the calibration from taking place
successfully.
If the problem continues, contact CELY or your supplier.
After calibration, the scales should be revised to check that calibration and linearity are correct. If necessary,
repeat calibration, and make sure that the scales are in a stable position before accepting any weight.
11. ERROR
CODES
ERROR
CODES
DESCRIPTION
SOLUTION
- - - - -
Above the range
Remove the weight from the scales.
If the problem continues, contact your distributor or CELY
for assistance.
Err 1
Date set-up error
Enter the date using the correct format and logical values.
Format: yy.mm.dd
Err 2
Time set-up error
Enter the time using the correct format and logical values.
Format: hh.mm.ss.
Err 4
Zero set-up error
The scales were not correctly set-up in the zero range, or
they were switched on when was pressed
.
Remove the weight from the scales and try again.
Use
to set the display to zero.
If the problem continues, contact your distributor or CELY
for assistance.
Err 6
A/D out of range
The A/D converter values are out of the normal range.
Remove the weight from the scales if there is excessive
weight and check that the platform is in position.
This error may indicate that the platform or electronics may
not be working correctly.