07 Maintenance
7.1 Cleaning
1. After printing, flush the microvalve, nozzle and Luer lock adapter to clean them and prevent clogging.
2. Disassemble the printhead (Section 4.2).
3. Mount the Luer lock adapter to a syringe loaded with deionized water or ethanol.
4. Flush the Luer lock adapter to remove residue. Let it air dry.
5. Connect a clean syringe to the provided pneumatic tubing.
6. Pour deionized water or ethanol into a plastic cup. If using water, use a mild temperature.
7. Retract the water/ethanol into the syringe (Figure 25A).
8. Connect the end of the tubing to the microvalve (Figure 25B).
9. Pressurize the syringe by pushing on the plunger and inserting the microvalve into the adapter tool. (Figure 25 CD). Fluid will flush the valve
when the valve opens (Figure 26).
Содержание 000000020566
Страница 1: ...BIO X EMD Printhead User Manual ...
Страница 2: ......
Страница 4: ......
Страница 5: ...1 Package contents 01 ...
Страница 7: ...3 Technical specifications 02 ...
Страница 11: ...7 Safety information 03 ...
Страница 13: ...9 03 Safety information Figure 2 Do not touch the outlined area of the EMD microvalve ...
Страница 15: ...11 Getting started 04 ...
Страница 40: ...36 04 Getting started Figure 23 Depressurizing and removing the EMD Printhead A B ...
Страница 43: ...39 Relevant G code commands 05 ...
Страница 46: ...42 ...
Страница 47: ...43 Frequently asked questions 06 ...
Страница 49: ...45 Maintenance 07 ...
Страница 51: ...47 07 Maintenance Figure 25 Preparing the microvalve for cleaning Figure 26 Flushing the microvalve A B C D ...
Страница 54: ...50 ...
Страница 55: ...51 Appendix A Consumables needles and nozzles ...
Страница 60: ...www cellink com ...