04 Getting started
3. The pressure can be adjusted within the cartridge to control the expulsion rate of the droplet.
• Certain bioinks require a minimum pressure to be expelled.
• A pressure that is too high can cause an expulsion rate that is too high, resulting in splattering of the droplet on the surface.
• A pressure that is too low might cause material to accumulate at the valve tip, or deflected droplets.
4. Use the Dispense button to test droplet expulsion and optimize your printing parameters.