04 Getting started
4.9 Printing optimization
The EMD Printhead uses an electromagnetically actuated valve to form droplets in the micro- and nanoliter range.
Pressurized fluid can flow when the valve is open, and the flow stops when it is closed. An optimal nozzle diameter,
valve travel speed and actuation yield the minimal dispensing volume. The BIO X allows the user to enter the open time
and the cycle time.
Due to the technique used to form the droplets, this printhead might be incompatible with highly viscous materials.
However, adding heat to thermosensitive materials can lower the viscosity enough to make them compatible. Larger
nozzle diameters and valve travel are recommended when using larger flow volumes and high-viscosity fluids, while
smaller diameters with smaller valve travel are recommended when using small dispensing volumes and low-viscosity
fluids.
Figure 24:
Diagram of valve
diameter and valve travel.
Содержание 000000020566
Страница 1: ...BIO X EMD Printhead User Manual ...
Страница 2: ......
Страница 4: ......
Страница 5: ...1 Package contents 01 ...
Страница 7: ...3 Technical specifications 02 ...
Страница 11: ...7 Safety information 03 ...
Страница 13: ...9 03 Safety information Figure 2 Do not touch the outlined area of the EMD microvalve ...
Страница 15: ...11 Getting started 04 ...
Страница 40: ...36 04 Getting started Figure 23 Depressurizing and removing the EMD Printhead A B ...
Страница 43: ...39 Relevant G code commands 05 ...
Страница 46: ...42 ...
Страница 47: ...43 Frequently asked questions 06 ...
Страница 49: ...45 Maintenance 07 ...
Страница 51: ...47 07 Maintenance Figure 25 Preparing the microvalve for cleaning Figure 26 Flushing the microvalve A B C D ...
Страница 54: ...50 ...
Страница 55: ...51 Appendix A Consumables needles and nozzles ...
Страница 60: ...www cellink com ...