PROBLEM
CAUSES
SOLUTIONS
SYSTEM WILL NOT TURN ON
System not plugged in and/or power
cord not properly seated in the
generator outlet on back of
instrument
Ensure power cord is plugged in and properly
seated with one end attached to the generator
and the other end to the power outlet
Power is not being delivered to outlet
After power is delivered to outlet, turn system
on
Fuse is missing or blown on system
Call Celleration Customer Service
NO POWER TO TREATMENT WAND
(NO SCREEN DISPLAY)
System is not turned on
Toggle power switch located on back of system
to ensure its in the “on” position
Treatment wand is not connected to
the generator
Insert or re-insert treatment wand cable into
connector or generator
SINGLE KEY ON KEYPAD NOT
FUNCTIONING
Mechanical failure
Call Celleration Customer Service
ALL KEYS ON KEYPAD NOT
FUNCTIONING
System software error
Shut down system by turning off power and
then restart. If error continues, call Celleration
Customer Service
OVERHEAT
Treatment wand temperature has
exceeded the allowed limit
The system alarm for overheating indicates
that the ultrasound is being turned off as a
preventive measure and does not indicate a
malfunction of the system
20 minutes is the expected maximum treatment
time after periods of system inactivity.
The overheat screen will automatically begin
to count down the minutes required to cool
the system. Once this completes, the “paused
treatment” screen will appear and the
treatment can now be completed.
DO NOT TURN OFF SYSTEM during cool
down(this will require a replacement of the
applicator)
ERROR MESSAGE ON DISPLAY
SCREEN
An error has occurred
Refer to the common errors table to determine
appropriate action. If error continues, note
the contents of the error message and call
Celleration Customer Service
TROUBLESHOOTING
Troubleshooting continued on next page
PL-61044_B
8
ULTRAMIST
™
SYSTEM
INSTRUCTIONS FOR USE