COMMON ERRORS (noted on display screen)

#

ERROR

CAUSE

SOLUTIONS

008

OVERHEAT

Treatment wand temperature has 

exceeded the allowed limit

The system alarm for overheating indicates that 

the ultrasound is being turned off as a preventive 

measure and does not indicate a malfunction of 

the system 
20 minutes is the expected maximum treatment 

time after periods of system inactivity.
The overheat screen will automatically begin 

to count down the minutes required to cool 

the system.  Once this completes, the “paused 

treatment” screen will appear and the treatment 

can now be completed.
DO NOT TURN OFF SYSTEM during cool down (this 

will require a replacement of the applicator)

200

INVALID 

REPLACE APPLICATOR

Applicator has been previously used 

and should be replaced

Replace applicator

201

REPOSITION

APPLICATOR

Applicator not positioned properly on 

treatment wand

Adjust current applicator properly on treatment 

wand

202

MAX USE

REPLACE APPLICATOR

Applicator has been used for the 

maximum time period allowed

Replace with new applicator to continue

203

NOT READING 

APPLICATOR

Applicator is not being sensed by 

treatment wand

Properly align the applicator on the treatment 

wand

For errors not listed or more details, refer to the Troubleshooting section of the Instructions for Use or contact Celleration 

Customer Service for futher information.

INDICATOR LIGHT (located on front of generator)

RED

A

n error has occurred - refer to the Troubleshooting section for more details

GREEN

Electrical/System Power On, System Ready and Functioning

PL-61044_B

       

7

ULTRAMIST

 

SYSTEM

  

INSTRUCTIONS FOR USE

Содержание Ultramist

Страница 1: ...Wand Tip APPLICATOR 18 Applicator 19 Tubing Cable Clip 20 Tubing Clamp 21 Infusion Tubing 22 Tubing Spike 23 Tubing Spike Protective Cover PLEASE READ INSTRUCTIONS BEFORE USE UltraMIST System ULTRAMIS...

Страница 2: ...ed performance DESCRIPTION INDICATIONS FOR USE Description The UltraMISTTM System delivers low frequency ultrasound to the treatment site using a noncontact saline mist Indications for Use The MIST Th...

Страница 3: ...the keypad to enter the size of each wound to be treated per the screen prompts 5 The total saline requirement for the combined treatment area is displayed approximately 100mL for every 25 cm2 Select...

Страница 4: ...d close the pump door Leave enough tubing to allow the saline bag to hang freely with the majority of the tubing between the pump and treatment wand 13 Open the tubing clamp Saline will not flow until...

Страница 5: ...or best results the treatment wand should be perpendicular to the wound and parallel to the floor during treatment Note It is important to maintain a 0 5 to 1 5 cm 0 2 0 6 inch distance from the leadi...

Страница 6: ...l data is encrypted to prevent usage by unauthorized personnel WARNING DO NOT ATTACH ANY UNAPPROVED USB DRIVES CABLES OR EQUIPMENT TO THE DRIVE AS IT MAY INTERFERE WITH OPERATION OF THE DEVICE Accessi...

Страница 7: ...PLICATOR Applicator has been previously used and should be replaced Replace applicator 201 REPOSITION APPLICATOR Applicator not positioned properly on treatment wand Adjust current applicator properly...

Страница 8: ...software error Shut down system by turning off power and then restart If error continues call Celleration Customer Service OVERHEAT Treatment wand temperature has exceeded the allowed limit The system...

Страница 9: ...r of applicator Ensure the applicator is positioned correctly with treatment wand tip located in center of applicator opening and fully seated on the applicator housing Tubing spike not fully mated wi...

Страница 10: ...Celleration Customer Service PHONE 1 866 307 MIST 6478 ADDRESS 6321 Bury Drive Suite 15 Eden Prairie MN 55346 WEBSITE www Celleration com EMAIL customerservice celleration com TECHNICAL SERVICE ULTRA...

Страница 11: ...re This End Up Keep Dry Recycle Authorized Representative in the European Community CE Mark Temperature Limit Consult Instructions for Use Manufacturer Date of Manufacture Lot Number Prescription Use...

Страница 12: ...ation of nearby equipment It may be necessary to take mitigation measures such as re orienting or relocating the UltraMIST System or shielding the location Harmonic emissions IEC 61000 3 2 Class A Vol...

Страница 13: ...stallation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful inference to radio communications Ho...

Отзывы: