PL-61044_B

       

12

The UltraMIST Therapy System is intended for use in the electromagnetic environment specified below. The customer or the user 

of the UltraMIST Therapy System should assure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF emissions

CISPR 11

Group 1

The UltraMIST System uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not likely to cause any 

interference in nearby electronic equipment.

RF emissions

CISPR 11

Class A

The UltraMIST System is suitable for use in all establishments other 

than domestic, and may be used in domestic establishments and those 

directly connected to the public low-voltage power supply network that 

supplies buildings used for domestic purposes, provided the following 

warning is heeded:

Warning:

 This equipment/system is intended for use by healthcare 

professionals only.  This equipment/system may cause radio 

interference or may disrupt the operation of nearby equipment.  It 

may be necessary to take mitigation measures, such as re-orienting or 

relocating the UltraMIST System or shielding the location.

Harmonic emissions

IEC 61000-3-2

Class A

Voltage fluctuations / flicker 

emissions

IEC 61000-3-3

Complies

Ref. IEC 60601-1-2:2007 §5.2.1.1

Emissions Test

Compliance

Radiation Guidance

Maximum Acoustic Output 

Power

Complies

The UltraMIST System requirement for the Maximum Acoustic Output 

Power is that the measured value be less than 3W/cm2, which is a 

requirement established by the World Health Organization for acoustic 

safety.  The actual output is significantly less than the above limit, which 

is considered to be safe. 

Sidewall Radiation

Complies

The UltraMIST System requirement for Sidewall Radiation is that the 

measured value be less than 100mW/cm2, which is a requirement 

established by the World Health Organization for acoustic safety.  

The actual output is significantly less than the above limit, which is 

considered to be safe.

Ref. TR-75143 on file at Celleration

GUIDANCE AND MANUFACTURER’S DECLARATION – RADIATION EMISSIONS

GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTRONIC EMISSIONS

ULTRAMIST

 

SYSTEM

  

INSTRUCTIONS FOR USE

Содержание Ultramist

Страница 1: ...Wand Tip APPLICATOR 18 Applicator 19 Tubing Cable Clip 20 Tubing Clamp 21 Infusion Tubing 22 Tubing Spike 23 Tubing Spike Protective Cover PLEASE READ INSTRUCTIONS BEFORE USE UltraMIST System ULTRAMIS...

Страница 2: ...ed performance DESCRIPTION INDICATIONS FOR USE Description The UltraMISTTM System delivers low frequency ultrasound to the treatment site using a noncontact saline mist Indications for Use The MIST Th...

Страница 3: ...the keypad to enter the size of each wound to be treated per the screen prompts 5 The total saline requirement for the combined treatment area is displayed approximately 100mL for every 25 cm2 Select...

Страница 4: ...d close the pump door Leave enough tubing to allow the saline bag to hang freely with the majority of the tubing between the pump and treatment wand 13 Open the tubing clamp Saline will not flow until...

Страница 5: ...or best results the treatment wand should be perpendicular to the wound and parallel to the floor during treatment Note It is important to maintain a 0 5 to 1 5 cm 0 2 0 6 inch distance from the leadi...

Страница 6: ...l data is encrypted to prevent usage by unauthorized personnel WARNING DO NOT ATTACH ANY UNAPPROVED USB DRIVES CABLES OR EQUIPMENT TO THE DRIVE AS IT MAY INTERFERE WITH OPERATION OF THE DEVICE Accessi...

Страница 7: ...PLICATOR Applicator has been previously used and should be replaced Replace applicator 201 REPOSITION APPLICATOR Applicator not positioned properly on treatment wand Adjust current applicator properly...

Страница 8: ...software error Shut down system by turning off power and then restart If error continues call Celleration Customer Service OVERHEAT Treatment wand temperature has exceeded the allowed limit The system...

Страница 9: ...r of applicator Ensure the applicator is positioned correctly with treatment wand tip located in center of applicator opening and fully seated on the applicator housing Tubing spike not fully mated wi...

Страница 10: ...Celleration Customer Service PHONE 1 866 307 MIST 6478 ADDRESS 6321 Bury Drive Suite 15 Eden Prairie MN 55346 WEBSITE www Celleration com EMAIL customerservice celleration com TECHNICAL SERVICE ULTRA...

Страница 11: ...re This End Up Keep Dry Recycle Authorized Representative in the European Community CE Mark Temperature Limit Consult Instructions for Use Manufacturer Date of Manufacture Lot Number Prescription Use...

Страница 12: ...ation of nearby equipment It may be necessary to take mitigation measures such as re orienting or relocating the UltraMIST System or shielding the location Harmonic emissions IEC 61000 3 2 Class A Vol...

Страница 13: ...stallation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful inference to radio communications Ho...

Отзывы: