![CEFLA MyRay Hyperion X9 Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/cefla/myray-hyperion-x9/myray-hyperion-x9_operators-manual_2578947061.webp)
EN
OPERATOR'S MANUAL
61
[11x5]
HiRes
kV
90
90
90
90
90
mA
6.00
6.00
6.00
6.00
6.00
exposure time (s)
9.0
9.0
9.0
9.0
9.0
mAs
54.00
54.00
54.00
54.00
54.00
CTDI100 [mGy]
7.748
6.647
6.944
6.468
6.607
[8x8]
**
kV
90
90
90
90
90
mA
4.00
4.00
4.00
4.00
4.00
exposure time (s)
3.6
3.6
3.6
3.6
3.6
mAs
14.40
14.40
14.40
14.40
14.40
CTDI100 [mGy]
2.080
1.519
1.664
1.448
1.600
[8x8]
HiRes
**
kV
90
90
90
90
90
mA
6.00
6.00
6.00
6.00
6.00
exposure time (s)
9.0
9.0
9.0
9.0
9.0
mAs
54.00
54.00
54.00
54.00
54.00
CTDI100 [mGy]
7.071
5.137
5.622
5.224
5.373
[8x5]
kV
90
90
90
90
90
mA
4.00
4.00
4.00
4.00
4.00
exposure time (s)
3.6
3.6
3.6
3.6
3.6
mAs
14.40
14.40
14.40
14.40
14.40
CTDI100 [mGy]
1.780
1.266
1.266
1.266
1.266
[8x5]
HiRes
kV
90
90
90
90
90
mA
6.00
6.00
6.00
6.00
6.00
exposure time (s)
9.0
9.0
9.0
9.0
9.0
mAs
54.00
54.00
54.00
54.00
54.00
CTDI100 [mGy]
5.902
4.204
4.204
4.204
4.204
[5x5]
kV
90
90
90
90
90
mA
5.00
5.00
5.00
5.00
5.00
exposure time (s)
3.6
3.6
3.6
3.6
3.6
mAs
18.00
18.00
18.00
18.00
18.00
CTDI100 [mGy]
2.421
1.520
1.550
1.437
1.502
[5x5]
HiRes
kV
90
90
90
90
90
mA
7.00
7.00
7.00
7.00
7.00
exposure time (s)
9.0
9.0
9.0
9.0
9.0
mAs
63.00
63.00
63.00
63.00
63.00
CTDI100 [mGy]
7.609
4.760
4.812
4.476
4.724
(*=available only with CBCT 1308 DXT detector, with Extended View software license)
(**=available only with CBCT 1313 DXT detector)
(***=available only with CBCT 1313 DXT detector, with Extended View software license)
Accuracy of declared values