background image

Instructions 

DPG2000B D4, M0, M1 Series

¼" NPT

2.88"

3.38"

0.75"

1.65"

Types of Gauges

Gauge reference reads zero with the gauge port open. 
Bipolar ranges read positive pressure and vacuum in the same 
units, and zero with the gauge port open.
Compound ranges read positive pressure in psig and vacuum 
in inHg, and zero with the gauge port open.
Sealed reference reads zero with the gauge port open and is 
referenced to 14.7 psi. Used for 1000 psi and up.
Absolute reference reads atmospheric pressure with gauge 
port open and zero at full vacuum. With the gauge port open 
to atmosphere, it is normal for readings to fluctuate due to 
continuously changing barometric pressure.

Power Up

Press and hold the front button for approximately 1 second.
The display is tested, the full-scale range is indicated, the 
display test is briefly shown again, then the actual pressure and 
units are displayed. The gauge is ready for use.

Power Up with Zero

This applies to gauge reference models only. Absolute refer-
ence gauges do not use the zero feature since they read 
atmospheric pressure under normal conditions.
Be sure the gauge port is exposed to normal atmospheric pres-
sure. The zeroing function is only stored at power-up and the 
zero correction is erased when the gauge is shut off.
Press and hold the front button until 

oooo

 is displayed.

Release the button. The gauge in now zeroed. The full-scale 
range is indicated and the display test is briefly shown again, 
then the actual pressure and units are displayed. The gauge 
is ready for use.
Attempting to zero the gauge with pressure greater than about 
3% of full-scale pressure or vacuum applied will result in an 
error condition, and the display will alternately indicate 

Err 0

 

and the actual pressure. The gauge must be powered down to 
reset the error condition. 

Normal Operation

Following the start-up initialization, the display indicates the 
pressure reading updated approximately 3 times per second. 
The auto shutoff timer starts when the gauge is powered up or 
whenever the button is pushed, unless the gauge shutoff time 
was set to zero for on/off operation. 
If excessive vacuum is applied to a pressure-only gauge, the 
display will indicate 

-Err

 until the vacuum is released. 

Applying vacuum to a gauge designed for pressure may 
damage the pressure sensor. If excessive pressure is applied 
(112.5% over range), an out-of-range indication of 

I – – –

 or 

I.–.–.–

 will be displayed depending on model.

Shut-Down

To shut off the gauge manually at any time, press and hold 
the button until the display indicates 

OFF

 (about 5 seconds) 

and then release.
When an auto shutoff timer is used, the display indicates 

OFF 

five seconds prior to auto shutoff. Press the button to keep the 
gauge on. The auto shutoff and backlight (if equipped) timers 
are reset whenever the button is pressed.

If the gauge is set up without auto shutoff (on/off operation) it will 
stay on until manually shut off or until the batteries are depleted. 
Turn gauge off when not in use to conserve battery life. 

Display Backlighting (BL Option Only)

Display backlighting will operate for one minute when the 
button is pressed provided the front light sensor detects low 
ambient light levels. 
Backlighting may not be apparent under some lighting condi-
tions. Backlighting may also be configured to be on at all times 
(which will shorten battery life), or be turned off at all times.

M0 Versions

M0 models are configured with minimum and maximum func-
tions disabled. One or both can be enabled or disabled in the 
User Configuration mode. 

Minimum/Maximum Memory

If enabled, minimum and maximum readings are stored con-
tinuously whenever the gauge is on. The stored readings can 
be manually cleared if desired. The MIN and MAX memory 
can be configured to save or clear the reading whenever the 
gauge is off.
Press and hold the button for about 1 second until 

MAX

 is 

displayed alternating with the units. The maximum reading will 
be continuously updated. The gauge may be left in this mode.
After 

MAX

 is displayed, press and hold the button for about 1 

second until 

MIN

 is displayed alternating with the units. The 

minimum reading will be continuously updated. The gauge 
may be left in this mode. If excessive vacuum is applied to a 
pressure-only gauge while in this mode, the display will indi-
cate 

-Err

 until the MAX/MIN readings are cleared. 

After 

MIN

 is displayed, press and hold the button again for 

about 1 second until 

* * * * *

 is displayed. The MAX and MIN 

memory is not erased and the gauge returns to normal opera-
tion with the display indicating the current reading.
Press and continue to hold the button until the display indicates 

clr MX/MN

 (about 3 seconds total) and then release the but-

ton. Both maximum and minimum values are cleared and the 
gauge returns to the normal operating mode. 

M1 Versions: Peak Reading

M1 models are peak reading gauges that only display and 
capture maximum readings. The maximum reading will be 
continuously updated. The MAX memory can be configured to 
save or clear the reading whenever the gauge is off. 

Battery Replacement

Precautions

A low battery indication (either LOBAT or a 

symbol depend-

ing on the model) will be shown in the upper left-hand corner 
of the display when the battery voltage falls sufficiently. The 
batteries should be replaced when the indicator comes on or 
unreliable readings may result. 
WARNING: Replace batteries with approved type in non-
hazardous locations only. Replace batteries with two Panasonic 
LR03 1.5 V AAA alkaline cells.
Replace both batteries with new ones at the same time. Do 
not mix different types of batteries. Substitution of components 
may impair intrinsic safety. 
1. Remove the 6 Phillips 

screws on the back of 
the unit. 

2.  Remove batteries by lift-

ing up the positive end of 
the battery (opposite the 
spring) taking care not to 
bend the spring.

3. Discard old batteries 

properly, do not discard 
into fire, sources of 
extreme heat, or in any 
hazardous manner.

4.  Install batteries with correct orientation. The negative (flat) 

end of each battery should be inserted first facing the bat-
tery holder spring.

5.  Replace the back cover, including the rubber gasket.

Approved Locations

The DPG2000B series is approved for use in the following 
Hazardous Locations.

IS Class I Div 1 Gp ABCD  

T3C Ta = –40ºC to 82ºC;  T4 Ta = –40ºC to 66ºC. 

CL I Zone 0 AEx/Ex ia IIC  

T3 Ta = –40ºC to 82ºC;  T4 Ta = –40ºC to 66ºC 

Installation 

4

 Read these instructions before installing the gauge. 

Configuration may be easier before the gauge is installed. 
Contact the factory for assistance.

4

 Installation instructions must be strictly followed in compli-

ance with Intrinsic Safety National Standard NEC 504 or 
ANSI/ISA RP 12.6 and the National Electrical Code. 

4

 Outdoor or wash down applications require a NEMA 4X 

gauge or installation in a NEMA 4X housing.

4

  Use fittings appropriate for the pressure range of the gauge. 

4

 Due to the hardness of stainless steel, it is recommended 

that a thread sealant be used to ensure leak-free operation.

4

 For contaminated media use an appropriate screen or filter 

to keep debris out of gauge port. 

4

 Avoid permanent sensor damage! NEVER insert objects into 

gauge port or blow out with compressed air. 

4

 Remove system pressures before removing or installing 

gauge.

4

 Install or remove gauge using a wrench on the hex fitting 

only. Do not attempt to turn by forcing the housing. 

Operation

4

 Use within the pressure range indicated on gauge label.

4

 Avoid permanent sensor damage! Do not apply vacuum to 

gauges not designated for vacuum operation.

4

 Use only with media compatible with 316L stainless steel. 

 Gauges are not for oxygen service. Accidental rupture of 

sensor diaphragm may cause silicone oil inside sensor to 
react with oxygen.

4

 The DPG2000B series gauges must only be operated in 

specified ambient temperature ranges.

Maintenance 

4

 The non-metallic cover of the pressure gauge is considered 

to constitute an electrostatic discharge hazard. Clean only 
with a damp cloth.

4

 Batteries must be replaced when the low battery indication 

comes on to prevent unreliable readings. 

4

 WARNING: Replace batteries with approved type in non-

hazardous locations only. 

4

  Approved batteries are 

t

wo Panasonic LR03 1.5 V AAA alka-

line cells. 

Replace both batteries at the same time.

WARNING: Substitution of batteries may impair intrinsic 
safety. 

Improper voltages will damage the gauge.

4

 WARNING: Substitution of components may impair intrinsic 

safety. Do not modify the gauge. 

4

 These products do not contain user-serviceable parts 

except for batteries. Contact factory for repairs, service, or 
refurbishment.

DS-DPG2000B rev. 12-12

Two 
AAA  

batteries

cecomp.com

1220 American Way Libertyville, IL 60048 

800-942-0315

CECOMP

 

Div. of

 

A

A

BSOLUTE

BSOLUTE

 P

 P

ROCESS

ROCESS

 I

 I

NSTRUMENTS

NSTRUMENTS

Cecomp maintains a constant effort to upgrade and improve its products. 

Specifications are subject to change without notice. See cecomp.com for 

latest product information. Consult factory for your specific requirements.

WARNING:

 This product can expose you to chemicals including 

nickel and chromium, which are known to the State of Califor-

nia to cause cancer or birth defects or other reproductive harm. 

For more information go to www.P65Warnings.ca.gov

Disposal of Electrical & Electronic Equipment 

Applicable 

in the European Union and other European countries. 
This product should not be treated as household waste 

when you wish to dispose of it. Please contact your local 

waste disposal authority for information and collection 

location for recycling electrical and electronic equipment. 

You may also return this product via pre-paid shipping 

to Absolute Process Instruments or your supplier for proper disposal.

Отзывы: