Sect. 3) Selecting the cycle from the 9 user (PERS) and 9 pre-configured (FIX) cycles

The controller features 9 custom cycles, called user cycles, and 9 pre-configured cycles, 
called fixed cycles.

Press the  

Se l e c t

P r o g .

  button until the PERS light turns on, if the desired cycle is in the user 

group.   Press   the   same   button   until   the   FIX   light   turns   on,   if   the   cycle   is   in   the   pre-

configured group. Then use the  

  and  

  buttons to select the cycle from the 9 

available possibilities (the number of the cycle is shown on the “PRG” display).

-----------------------------------------------------------------------------------------

Example:

To select cycle number 4 in the user group.

Make sure the “PERS” light is on, if not press 

Se l e c t

P r o g .

. Then, press the 

 button until the 

number 4 appears on the “PRG” display.

-----------------------------------------------------------------------------------------

Sect. 4) Entering or modifying the selected cycle

Press the  

C h a n g e

  button for at least half a second. An audible signal will indicate that the 

modify cycle procedure has been activated, and the main display will show the time of the 
first step, flashing. If the cycle is empty (that is, has not yet been entered), the time shown 
will be 00.00, otherwise the previously set time in hours:minutes will be displayed (e.g. 
02.25 indicates a time of 2 hours and 25 minutes).

Use the 

and 

 buttons to modify the value, if required, or press 

En t e r

 to move to 

the next entry, that is, the setpoint. If the cycle is empty, the display will show a minimum 
setpoint equal to the ambient temperature (e.g. 25, that is, 25°C), otherwise it will show the 
previous setpoint.

Again use the 

and 

buttons to modify the value, if required, or press 

En t e r

 again to 

move to the time/setpoint pair for the following step. 
The display (STEP) will show which step is being displayed and/or modified. By setting a 
time equal to 00.00, the cycle will end at the step in question. To end the modify cycle 

procedure without scrolling the entire cycle, press the 

C h a n g e

 button again. 

NOTE

During this procedure, the time or setpoint values displayed will flash on the display 
(MAIN).
The new cycles entered or the modifications made to the previously entered cycles in 
the user cycle group are stored permanently, even when the controller is off.

______________________________________________________________________________________

6

Содержание MicroProgel TC 32

Страница 1: ...ammarby Fabriksv g 43 S 120 33 STOCKHOLM Tel 46 8 658 50 04 Fax 46 8 669 60 98 G teborg Argongatan 8 S 431 53 M LNDAL Tel 46 31 87 51 30 Fax 46 31 87 97 61 Web www cebex se Mail Info cebex se Manual M...

Страница 2: ...user PERS and 9 pre configured FIX cycles 6 Sect 4 Entering or modifying the selected cycle 6 Sect 5 Running a cycle 8 Sect 6 Activating the timer for the delayed running of the cycle 8 Sect 7 Settin...

Страница 3: ...cycle belongs to the user group customisable The FIX light indicates that the cycle belongs to the group of cycles pre configured by the installer Display STEP of the step in progress If the controll...

Страница 4: ...G and STEP displays indicate the cycle and step in progress The group of lights indicate whether in the current step the temperature is increasing heating stabilising or decreasing cooling The energy...

Страница 5: ...e the last time the counter was reset by the user the kWh light is on In START mode Normally the controller displays the temperature of the kiln in C on the MAIN display and the C light is on Pressing...

Страница 6: ...ep flashing If the cycle is empty that is has not yet been entered the time shown will be 00 00 otherwise the previously set time in hours minutes will be displayed e g 02 25 indicates a time of 2 hou...

Страница 7: ...layed is 01 00 that is 1h flashing press En te r the controller prompts the setpoint for the first time setpoint pair STEP no 1 press to increase the value displayed or to decrease it until the temper...

Страница 8: ...d to display the following values in sequence Operating setpoint Time elapsed since the start Partial energy consumed since the last start Total energy consumed this value can be reset see the corresp...

Страница 9: ...ing for the end of the delay time by pressing the S T O P S T A R T button The time displayed can be increased or decreased using the and buttons WARNING changes made in this way to the start delay ti...

Страница 10: ...er ER31 Hardware fault Contact the installer ER41 ER42 Thermocouple continuity failure not connected or connected incorrectly Check the thermocouple and connections ER43 Cold junction temperature sens...