12
RS-232 Serial Connection
The Remote Display supports an RS-232 connection to data consumers, such as a computer
or PLC. To connect the Remote Display to an external RS-232 device, use the following
procedure:
1. Find the female RS-232 connector on the rear panel of the Remote Display Mux Box.
2. Using a shielded RS-232 cable, connect the Mux Box to the external data consumer.
Depending on the device and cable configuration, it may be necessary to use a
female-to-male gender changer. They can be used freely without any impact on the
Remote Display functionality.
3. Use the following RS-232 settings on the external device to establish communication
with the Remote Display Mux Box.
Speed (baud)
Data Bits
Bits Stop
Bits Parity
Flow Control
9600
8
1
None
XON/XOFF
4. Refer to the Communication Protocol section below for a list of available commands.
Many modern computers use USB ports in lieu of RS-232 ports. The Remote Display
can still communicate to these devices over RS-232 by using a Serial-to-USB converter.
There are many such devices available. When searching, ensure that the device is
compatible with the Operating System version that is being used on the computer.
Communication Protocol
The following list summarizes the serial commands available for exchanging data with the
Remote Display Mux Box. These commands apply to both the RS-232 and PC USB
communication modes. Type:
“Enter”
or “Return”
after command.
Command
Description
A
Print current gage readings for all channels
N :A
Print current gage reading for Channel “N”. ex: 1:A for channel 1
D
Print data packet* for all channels
N :D
Print data packet* for channel “N”
R
Request a measurement update on all Digimatic channels on Tablet
N :R
Request a measurement update for channel “N” on Tablet
C
Clear gage measurements for all channels (quadrature only)
N :C
Clear gage measurement for channel “N” (quadrature only)
rstErr
Resets global errors
N :pXXX=?
Get the value of configuration parameter “XXX” for channel “N”
N :pXXX=<new value>
Set the value of parameter “XXX” for channel “N” to “<new value
х
”
Save
Saves any modified parameters to retentive flash
Help
Show help text and command list
* Data Packet format: Channel: Reading: Analog Scale: New Data: Max: Min
Note that the Analog Scale field has a maximum value of 1000.
Note N = channel number
Содержание EMS 4
Страница 1: ...1 Remote Display User Manual ...
Страница 10: ...10 ...
Страница 17: ...17 Find the preferred cloud storage service and follow the instructions to save the logs ...
Страница 20: ...20 ...
Страница 21: ...21 Digimatic Data Format ...
Страница 26: ...26 6 Select Advanced Options from the Troubleshoot menu 7 Select Startup Settings ...