E
E
P
P
by
29
En
Measurements
Continuity measurement
• Measurement range:
0
Ω
to 120
Ω
.
• Measurement current:
-
200 mA
if R
≤
2
Ω
,
20 mA
otherwise.
-
20 mA
(eco mode enabled) for R from 0
Ω
to 120
Ω
.
• Programmable conformity threshold (beep):
-
1
Ω
to
2
Ω
.
• Standard:
CEI 61557-4
.
Voltage alarm
Insulation measurement
• Range:
10 k
Ω
to
999M
Ω
.
• Configurable voltages
-
50 V DC
,
100 V DC
,
250 V DC
,
500 V DC
,
1 000 V DC
.
• Rated measurement current:
≥
1mA
.
• Adjustable conformity threshold:
OFF, 0.25 M
Ω,
0,5 M
Ω,
1 M
Ω,
2 M
Ω,
100 M
Ω.
•Standards:
CEI 61557-2
.
Resistance range
0.01
Ω
to 4.00
Ω
4.1
Ω
to 120.0
Ω
Resolution
0
.
01
Ω
0
.
1
Ω
Accuracy
±5% ±5 digits
±5% ± 5 digits
Resistance range
± 20 V to 1 500 V DC
± 20 V to 400 V 50/60 Hz AC
Resolution
1 V
2 V
Accuracy
5% ±5 digits
5% ± 5 digits
Resistance range
10 k
Ω
à 99.99 k
Ω
100 k
Ω
à 999.9 k
Ω
1 M
Ω
à 99.99 M
Ω
100 M
Ω
à 999.9 M
Ω
Resolution
0.01 k
Ω
0.1 k
Ω
0.01 M
Ω
0.1 M
Ω
Accuracy
±5% ±5 digits
±5% ± 5digits
±5% ±5 digits
±5% ± 5digits
M
E
N
U
TE
ST
CP
-10
0B
Long press
on TEST
waRning
The readings may be biased due to the impedance of additional parallel
circuits or by transient currents.