12
9
cod: ed. 09-2021
EN
G
LI
SH
It is advisable to install in parallel only identical
machines, having the same flow rate and the same
head. Two or three suction units in parallel* double
or triple the flow rate, provided that the diameter
of the main piping is increased of ten mm for each
additional suction unit added. In the same manner,
the diameter of the exhausted air piping must be
increased. Turbo- Smart HP always includes all
the accessories required for the installation in
parallel, therefore check valves, additional boards
or peripherals are not needed. Connected to the
same main piping, the two or three suction units in
parallel, shall be connected in parallel also to the
permissive
(12)
incoming from the various dental
units. While installing the suction units in parallel,
make sure not to invert the cables of no. 1 clamps
with no. 2 clamps. With suction units in parallel
it can occur that one unit stops without the operators
noticing it. To address this issue, the control unit
(see wiring diagram page 31) clamps
(dry contacts) carry an alarm signal at remote
distance. The Turbo- Smart HP units in parallel
work better and offer greater energy saving, when
they are active simultaneously, independently from
the request. Shutting the power supply to one of the
suction units does not result in energy and reduces
the performance of the units still in operation.
PARALLEL INSTALLATION
Содержание TURBO SMART HP
Страница 1: ...WE LOVE WHAT WE DO English Instruction Manual TURBO SMART HP...
Страница 2: ...2...
Страница 30: ...30 cod ed 09 2021 SAME LEVEL INSTALLATION LEVEL BELOW INSTALLATION ASSEMBLY SCHEME...
Страница 31: ...31 cod ed 09 2021 EN G LISH TURBO SMART HP CIRCUIT Fig C ELECTRICAL CONNECTIONS...
Страница 33: ...33 cod ed 09 2021 EN G LISH EXAMPLES OF PIPING Fig E E TURBO SMART HP Fig F...
Страница 34: ...34 cod ed 09 2021 Fig G TURBO SMART HP EXPLODED VIEW...
Страница 35: ...35 cod ed 09 2021 EN G LISH Fig H CENTRIFUGAL SEPARATOR 023764 EXPLODED VIEW...
Страница 36: ...36 cod ed 09 2021 Fig I UNI JET 75 SX 015663 EXPLODED VIEW...
Страница 37: ...37 cod ed 09 2021 EN G LISH Fig L UNI JET 75 DX 015664 EXPLODED VIEW...
Страница 38: ...38 cod ed 09 2021 Fig M AMALGAM SEPARATOR 043396 EXPLODED VIEW...
Страница 39: ...39 cod ed 09 2021 EN G LISH Fig N CONTROL UNIT 023764 EXPLODED VIEW...
Страница 40: ...40 cod ed 09 2021 Fig O TURBO SMART HP DIMENSIONS...
Страница 41: ...41 cod ed 09 2021 EN G LISH Fig P TURBO SMART HP DIMENSIONS...
Страница 42: ...42 cod ed 09 2021 TURBO SMART HP DIMENSIONS Fig Q...
Страница 43: ...43 cod ed 09 2021 EN G LISH...