Catalyst Enterprises Inc.
Protocol Analyzer Projects
28
Project Settings
To set project options click the
Settings
tab.
Figure 18. Setting Project Options
Trace Memory Status
Memory Size
To reduce the capture memory size, check Partial Memory and enter the
desired buffer size or, check entire memory to allow capture for the entire
memory (1 GB) if you want to capture the maximum amount of trace
data.
Note:
In cases where the size of a data packet exceeds the set buffer
memory allocation, the project will run, but no capture will result.
In such cases you must increase the buffer memory size to a value
greater than the packet size.
Trigger Position
Pre-Trigger is set to define the percentage of data to be captured before
and after the triggering event. It is set by default to 1% in the Snapshot
mode and to 50% in the Manual trigger and Pattern mode. You may
change this percentage by dragging the slider to the desired value.
The capture of the specified percentage of the data prior to the triggering
event cannot be guaranteed and may in some cases be 0. This can occur in
cases where the triggering event occurs before the required number of
pre-trigger event data can be stored. In these cases the data display will
show fewer than the specified data points prior to the triggering event. For
more detail see
Sample File Name
Click the ellipses next to the Sample File Name text box and choose a file
name and location for the results of your current project.
Auto Run
To repeat the current capture and trigger setup automatically, check the
Auto Run
checkbox and enter the number of times in the
Number to
Run
text box. The capture and trigger will repeat automatically for the
specified number of times and the results saved in consecutively
numbered
User Defined Sample File Name.was
files.
Содержание AirCom UWB Analyzer
Страница 1: ...March 2007 P N 400 0103 001 User s Manual Version 1 22 ...
Страница 4: ...Catalyst Enterprises Inc Copyright iv ...
Страница 54: ...Catalyst Enterprises Inc Faded Packets 46 Figure 32 Illustration of Faded Packets ...
Страница 76: ...Catalyst Enterprises Inc Display Configuration 68 ...
Страница 109: ......
Страница 110: ...LeCroy Corporation 3385 Scott Blvd Santa Clara CA 95054 Tel 1 800 909 7112 Fax 408 727 6622 www lecroy com ...