Oximeter Classifications
Electrical Insulation
The Oximeter (with integrated AC power supply) is a Class I device.
IEC 60601-1
Warning:
The Oximeter is not “Category AP or APG Equipment.”
There is a risk of an explosion hazard if a damaged Oximeter or accessory is
used in the presence of a flammable anesthetic mixture with air or with oxygen
or nitrous oxide.
Note:
The Oximeter can remain connected to the patient during cardiac defibrillation.
All applied parts are “Type BF Defibrillation Proof.”
Patient Environment
The Oximeter has been tested with specific parts of the system used within the patient
environment (Figure 1). These parts are:
•
The Oximeter
•
Appropriate accessories as listed in Appendix F,
Parts
on page 93
•
AC power cord
Figure 1:
Patient Environment
xvi
CAS Medical Systems, Inc.
Preface: Safety