INSTALLATION
P
OWER
S
OURCE
R
EQUIREMENTS
When selecting a location for the oven, verify
each
of the following requirements is satisfied.
Power Source
: The power supply must meet the power requirements listed on the oven data plate
(located on the back of the unit, beneath the power feed inlet.
•
These ovens are intended for a
220 – 240 volt, 50/60 Hz
applications at
10 amps
.
•
The power source must be
single (1) phase
and
protective earth grounded
.
•
The power source must conform to all national and local electrical codes.
•
Supplied voltage must not vary more than 10% from the data plate rating
. Damage to the
oven may result if the supplied voltage varies more than 10%.
•
The recommended wall circuit breakers for these units are 20 amps.
•
Use a separate circuit to prevent loss of the unit due to overloading or circuit failure. The
circuit must meet or exceed the amperage requirements listed on the oven data plate.
Power Cord.
The unit must be positioned so that all operators can quickly unplug the oven in the
event of an emergency.
•
The unit comes provided with a
250V, 10 Amp, 2.5m (8ft) Euro CEE7/7
power cord.
•
This unit is also provided with a
250V, 15A, 8ft 2in (2.5m) NEMA 6-15P
power cord.
Fuses
: The unit ships with two 10 Amp fuses. One in the power cord inlet and the other in an
adjacent fuse holder.
•
The fuses must be installed and intact for the unit to function.
•
Always determine the cause of a blown fuse before restoring the unit to operation.
EURO CEE7/7
Outlet
NEMA 6-15R
Outlet
Содержание CDO-5-2
Страница 1: ...Installation Operation Manual CDO 5 2 Drying Decarboxylation Oven 220 240 Voltage...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Страница 10: ...10 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Страница 44: ......