- 18 -
COUNT MODE(
Ⅱ
)
INITIAL DISPLAY
DESCRIPTION
Counting Mode
(
) Press the [TARE] key long to set the Count Mo
Ⅱ
de in detail
DISPLAY
KEY
DESCRIPTION
ZERO
Input the weight of unit with sample,
and go to count weighing mode
TARE
Input the weight of unit with key value,
and go to count weighing mode
End
ENTER Go to the count weighing mode
(
Ⅱ
- 1) In the case of pressing the [ZERO] key
DISPLAY
KEY
DESCRIPTION
SAMPLE
Sample weight input state.
UnLoAd
Unload
display
A/D (24750)
ENTER
Display unload value
LoAd
A/D (24750)
ENTER
After put a sample weight on the platform,
press [ENTER] key
Good /nUMbEr
10 pcs
After checking the message “Good” “nUMbEr”
10
Input the number what you want
(Ex) 10kg(sample) 10 , weight of unit is 1kg
Good
ENTER
After saving, go to initial display of count mode
ENTER It is displayed and then go to counter mode
(
Ⅱ
- 2) In the case of pressing the [TARE] key
DISPLAY
KEY
DESCRIPTION
Input the weight of sample with key
For example, set the 0.5kg to count ‘1’
WEiGht
0.500 kg
ENTER
After saving, go to initial display of count mode
If the weight of sample is over the maximum
Err 08 Err 32
Capacity(Err08) or small(Err32), return
ENTER It is displayed and then go to counter mode
SAMP = Z
In = t
on