GRAVITY COMPENSATION & THE OTHERS SET MODE
This mode set gravity compensation, dual range hold/unit conversion & decimal
specification.
1. How to Enter
Turn on while pressing CAL switch on the back side.
Press LIGHT/PRINT key and it will be moved to this mode.
2. Keys Usage
key: Increase the first place value by 1.
key: Used to set the current value
10.
key: Used to be moved to the next menu.
ZERO/MODE key: Used to set the current value to zero.
3. Description
Gravity Compensation Function:
This function allows it to be calibrated in one location and then adjusted to match
the acceleration of gravity at another location where it will be used.
Multi-interval(Dual range):
EX 1) If you calibrated minimum division to 1kg and maximum capacity to
3000kg
0
1500kg - minimum division is 0.5kg.
1500
3000kg - minimum division is 1kg.
EX 2) If you calibrated minimum division to 2kg and maximum capacity to
5000kg
0
2500kg - minimum division is 1kg.
2500
5000kg - minimum division is 2kg.
EX 3) If you calibrated minimum division to 5kg and maximum capacity to
10000kg.
0
4000kg - minimum division is 2kg.
4000
10000kg - minimum division is 5kg.
HOLD/UNIT conversion:
Select whether HOLD or UNIT conversion.
Decimal point:
Select whether ‘ . ’ or ‘ , ’