CAL 2
Function: Minimum Division Set
Range
0.001
50kg
Key Display
Description
Key :
Control division
Key :
Control digit
Key :
Store and move to
the next menu
CAL 2 condition.
Existing minimum division
Setting minimum division
Press key
Key Display
Description
Key :
Control division
Key :
Control digit
Key :
Store and move to
the next menu
CAL 3 condition.
100kg
10000kg
Press key
Note 1. The minimum division means the value of one division.
Note 2. External resolution is obtained by dividing the min. division
by the maximum capacity. Set the resolution to be within 1 / 10,000.
Note 3. The maximum capacity / minimum division is over 10,000, Error
message (“CH 11”) will occur.
Note 4. If you press key, it will be moved to CAL 3.
CAL 3
Function: Setting Weight in Span Calibration
Range
1
Maximum Capacity of CAL 1
Maximum capacity
value of CAL 1.
Setting maximum
capacity value
ex)
Note 1. The setting weight shall be within the range of 10%
100% of maximum
weight.
Note 2. If the setting weight over the maximum capacity or under the 10% of the
maximum weight, Error message (“CH 12”) will occur.