20
CAL 3
CAL 3-1
Function : Setting Multi Calibration Step
Range of set value : 1 ~ 5
KEY
DISPLAY
DESCRIPTION
▲,
▼
Key : Value change
ZERO Key : Cancel
SET Key : Save and next
menu
STEP- 1
Setting multi calibration for step 1
(CAL3-3 and CAL 3-4 are carried out once)
STEP- 5
Setting multi calibration for step 5
(CAL3-3 and CAL 3-4 are carried out five times.)
► Note 1. A function to use, when the output of load cell is corrected by setting multiple
points in some sections because the actual curve of load cell is not a straight line.
CAL 3-2
Function: Zero Calibration
KEY
DISPLAY
DESCRIPTION
ZERO Key : Cancel
SET Key : Save and next
menu
UnLoAd
Empty the load tray and press the setup key.
1234
The current weight value is displayed.
Confirm 'Stable' and press the setup key.
- - -
Zeroing in progress...
► Note 1. If zeroing finished with no error, it moves to Setting Weight (CAL 3-3) although
no key is pressed.