SERVICE SCHEDULE 6.1
Class II Microbiological Safety Cabinets
Schedule of work included in service agreement
1
Filter Integrity Test (D.O.P) on all main filter(s) and seals
2
Airflow Profiling ensuring compliance to latest British & European Standards
3
Operator Protection Testing (K.I. Discus) 1 x Per Annum
4
Check and adjust alarm parameters & control functions as required
5
Check and adjust electrical and electronic controls as related to the airflow system
6
Check and adjust operation of exhaust system as required
7
Mechanically inspect cabinet components and duct connections (if fitted)
8
Change main filters as necessary (Supply and fitting subject to additional charge)
9
Prepare computer generated Service Report
Above work to be carried out twice each service year
(Operator Protection Test (K.I. Discus) carried out once per year unless instructed otherwise
Service schedules can be tailored to suit individual needs
For a competitive Quotation please contact our service
Department on