11
5.3 d
IsInfeCtIon
p
roCess
!
Always operate the
UV-C LED BOXER
in a clean and dry location. Do not expose to
rain, humidity, saline water, sand or dust.
• Connect the power cable to the power connector
❺
.
• Connect the power supply to the main power or to a battery.
• Select ON on the main switch
❻
.
• To optimize the UVC LED efficiency let them run for a preliminary 5-minute cycle with no
equipment in the compartment, this will bring the transformers to the optimum operating tem-
perature.
• Open the upper lid with the 2 portholes. Place the objects to be disinfected on the grid. Do
not exceed the weight of 20 kg – 45 lbs.
• Place the Dosimeter card on the grid near the equipment
• Close the disinfection compartment repositioning the upper lid
⓬
Magnets will help to keep
the closed position.
• Set the time on the timer (as per 4.3), for COVID disinfection the recommended exposure of
22mJ/cm
2
can be achieved in 5 minutes.
• Activate the on/off switch
❹
, the red indicator will light up.
• Wait until the completion of the disinfection cycle, the red Led indicator will switch off.
• In case the compartment is opened before the completion of the disinfection cycle the LED
will switch off and the timer will reset to zero.
!
To ensure the correct disinfection process, the cycle of radiation has to be continuous.
In case of any interruption the cycle should be repeated entirely.
WARNING!
WARNING!
WARNING!
WARNING!