Humidity is higher than set point.
Is the set point within the unit specification range? Is the drain line connected
and draining properly? No kinks or up-hill traps in the line?
Is the internal air saturated? Turn humidity enable switch off, open door, and
set temperature to 100ºC for 1 hour to dry out the chamber.
Is water present in exhaust vent? Remove stopper and let water drain out.
Interior lights won’t come ‘on’
Are lights turned ‘on’?
Is the connector on the bottom of the door switch connected (see below)?
Glass fogs up on inside
Is the humidity switch on?
Is the connector on the bottom of the door switch connected (see picture)?
Is the door sealed tightly?
Is the door gasket collapsed?
Has the unit stabilized 1 hour?
Door
connector