36 Chapter 5 Acquiring Images using the Cephalometric Modality
5
On the control panel, press and hold
to level the ear cones to the patient’s auditory canals.
6
Gently push in at the
top part
of a head clamp bar to retract both head clamp bars and to fit the
ear cones into the ear auditory canals.
7
Lower the nasion support to a vertical position.
Launching the X-ray
To launch the X-ray, follow these steps:
1
Leave the X-ray room and close the door. You must keep visual contact with the patient during
acquisition.
Important: If you have a problem that requires you to stop the
acquisition, release the exposure button of the remote control or
press the red emergency stop button.