Cephalometric Modality User guide for CS 8100 Family and CS 8100 3D Family_SM987_Ed04
19
4
Getting Started
Switching on the Unit
Before switching on the unit, check that:
The installation of the unit is complete.
The computer is switched on.
To switch on the unit, follow these steps:
1.
On the unit column, press the ON button.
2.
You must wait for a minute for the connection between the unit and the computer to be
established.
If you start the Acquisition interface before the connection is established an error message is
displayed. Click OK on the error pop-up window. The Acquisition interface will close.
3.
You can now proceed to start the Acquisition interface from the imaging software.
Increasing the operating life of the X-ray tube
To increase the operating life of the X-ray tube, follow these steps:
1.
In the Panoramic Acquisition Window, select the Parameter pane.
2.
Select the following parameter setting: 70 kV - 6.3 mA
3
Leave the X-ray room and close the door.
4
Launch an X-ray for the parameter setting by pressing and holding the X-ray remote control
exposure button.
5
Repeat Steps 2 to 4 for the following parameters:
80 kV - 10 mA
85 kV - 10 mA
The unit is now ready to be used for acquisition.
Important: You must carry out the following procedures at the first
loading of the unit or if the unit has not been used for a month to
increase the operating life of the X-ray tube.
Содержание CS 8100 Access
Страница 6: ...2 Chapter 1 Conventions in This Guide...
Страница 26: ...22 Chapter 4 Getting Started...
Страница 52: ...48 Chapter 6 Maintenance 3 Select the desired test and follow the on screen instructions...
Страница 54: ...50 Chapter 7 Troubleshooting...
Страница 56: ...SM987 Ed04 2019 02 For more information visit www carestreamdental com...