English
9
ELECTROMAGNETIC COMPATIBILITY (EMC)
WARNING:
use of portable phones or other radio frequency (RF)
emitting equipment near the system may cause unexpected or
adverse operation
The Pulmolife has been tested to EN60601-1-2:2007, regarding its ability
to operate in an environment containing other electrical/electronic
equipment (including other medical devices).
The purpose of this testing is to ensure that the Pulmolife is not likely to
adversely affect the normal operation of other such equipment and that
other such equipment is not likely to adversely affect the normal operation
of the Pulmolife.
Despite the testing of the Pulmolife that has been undertaken, normal
operation of the Pulmolife can be affected by other electrical/electronic
equipment and portable and mobile RF communications equipment.
As the Pulmolife is medical equipment, special precautions are needed
regarding EMC (electromagnetic compatibility).
It is important that the Pulmolife is configured and installed/put into
service, in accordance with the instructions/guidance provided herein and
is used only in the configuration as supplied.
Changes or modifications to the Pulmolife may result in increased
emissions or decreased immunity of the Pulmolife in relation to EMC
performance.
The Pulmolife should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use with other equipment is necessary,
the Pulmolife and the other equipment should be observed/ monitored, to
verify normal operation in the configuration in which it will be used.
For the purposes of EN60601-1-2 the Pulmolife has an essential
performance (the Pulmolife must not begin to measure unintentionally
during a test and should pass a post test 3 litre calibration syringe
measurement).
Содержание PulmoLife
Страница 2: ......
Страница 4: ......
Страница 44: ...40 Deutsch 5 Sekunden selbstst ndig ab...
Страница 119: ...115 VII PULMOLIFE PulmoLife 1 1 1 1 1 2 PULMOLIFE...
Страница 120: ...116 WEEE IPX0 PulmoLife 35...
Страница 121: ...117...
Страница 122: ...118 ECCS NHANES III 45 ECCS NHANES III 0 00 PulmoLife 0 00 1 1...
Страница 123: ...119 3 1 80 1 20 90 30 90 1 100 3 CareFusion PeraSafe 36 SSC5000A PulmoLife 36 SST1000...
Страница 124: ...120 PeraSafe 10 PulmoLife PulmoLife 3 36 SM2125 Ch 0 00 3 3 3 2 91 3 09...
Страница 125: ...121...
Страница 126: ...122 MicroGard PulmoLife 3 bAt 5 3...
Страница 127: ...123 Prd 1 ECCS4 2 NHANES III5 3 6 rE PulmoLife bAt 5...
Страница 128: ...124 IEC 60601 1 B 3 CR 2450 IEC 60601 1 B...
Страница 133: ...129 PulmoLife PulmoLife IEC 60601 a b 1 80 800 2 a AM FM PulmoLife b 150 80 3 PulmoLife PulmoLife PulmoLife PulmoLife...
Страница 137: ...133 0086 93 42 WEEE SGS Testing Services OSHA...
Страница 138: ...134 Enter IX PULMOLIFE PulmoLife FEV1 FEV1 FEV1 FEV1 COPD1 2 PULMOLIFE...
Страница 139: ...135 WEEE IPX0 PulmoLife COPD COPD 35 COPD...
Страница 140: ...136 Enter ECCS NHANES III 45 ECCS NHANES III 0 00 PulmoLife 0 00 FEV1 FEV1...
Страница 141: ...137 3 FEV1 80 FEV1 20 90 30 90 FEV1 100 Enter CareFusion PeraSafe 36 SSC5000A PulmoLfe 36 SST1000 PulmoLife PeraSafe 10...
Страница 142: ...138 PulmoLife Pulmolife 3 36 SM2125 Ch 0 00 3 3 3 2 91 3 09 MicroGard...
Страница 143: ...139 PulmoLife 3V bAt 5 3V EU...
Страница 144: ...140 Enter Prd 1 ECCS4 2 NHANES III5 3 6 Enter rE Pulmolife bAt 5...
Страница 145: ...141 IEC 60601 1 B 3V CR 2450 IEC 60601 1 B...
Страница 153: ...149 B 0086 93 42 EEC WEEE Rx SGS...
Страница 155: ......