64
per minute BTPS depending upon the configuration in the PC
software.
The calibration should remain stable indefinitely, unless the
transducer is physically damaged, and the unit should not require
re-calibration.
However, as a check on the correct functioning of the unit we do
recommend that the calibration is checked periodically.
To check the calibration of the unit a large syringe should be used.
This should have a 30mm outlet and be of at least three litres in
capacity and preferably six litres.
A 3 Litre Syringe is available from CareFusion. The MicroDL should
be connected to this syringe with the minimum of adapters and the
syringe volume should be injected into the transducer evenly,
without pausing.
If, after checking, the unit appears to require re-calibration then
please refer to your supplier for service.
Looking after your MicroDL Spirometer
Please observe the following precautions:
*
Avoid exposing the MicroDL to direct sunlight.
*
Avoid operating the spirometer in dusty conditions or near to
heating appliances or radiators.
*
Do not keep the spirometer in a damp place or expose it to
extremes of temperature.
*
Do not direct the transducer holder towards a strong light
source whilst operating the spirometer.
Содержание MicroDL
Страница 1: ...1 MicroDL Operating Manual English Bedienungsanleitung Deutsch...
Страница 2: ......
Страница 4: ...2 IV 103 104 MicroDL 106 MicroDL SPCS 129 131 131 132 134...
Страница 36: ......
Страница 40: ...37 MicroDL 1 Fig 1 of enter 3 4 on 2...
Страница 73: ...70 MicroDL 1 Fig 1 of enter 3 4 on 2...
Страница 106: ......
Страница 107: ...103 MicroDL IV MicroDL CareFusion MicroDL Spirometry PC Software SPCS...
Страница 108: ...104 SPCS 200 CareFusion B T P S MicroDL 1 MicroDL 2 CareFusion 3 3 4 SPCS...
Страница 109: ...105 MicroDL 1 Fig 1 of enter 3 4 on 2...
Страница 110: ...106 MicroDL MicroDL SPCS FEV1 FVC PEF FEV1 FVC SPCS MicroDL ON OFF...
Страница 111: ...107 SPCS ENTER 1 ENTER...
Страница 112: ...108 MicroDL ENTER ENTER...
Страница 113: ...109 SPCS ENTER...
Страница 114: ...110 1 1999 ENTER ENTER...
Страница 115: ...111 ENTER ENTER ENTER MicroDL...
Страница 116: ...112 PEF FEV1 FVC SPCS...
Страница 117: ...113 PEF PEF SPCS SPCS PEF FEV1 FVC PEF FEV1 FVC...
Страница 118: ...114 enter off on enter off on PEF FEV1 on off enter FVC ENTER SPCS FEV1 FVC PEF FEV1 FVC 5 SPCS 10 1 10 4 SPCS ENTER...
Страница 119: ...115 ENTER EO E1 E4 ENTER ENTER...
Страница 120: ...116 2 ENTER 1 1999 ENTER...
Страница 121: ...117 ENTER...
Страница 122: ...118 ENTER...
Страница 123: ...119 PEF FEV1 FVC FEV1 FVC FEV1FVC SPCS ENTER OFF SPCS FEV1 FVC PEF FEV1 FVC OFF...
Страница 126: ...122 1 MicroDL...
Страница 128: ...124 FEV1 FVC PEF 2 ENTER 2 MicroDL 2 1 1 ENTER...
Страница 131: ...127 3 MicroDL 3 65 65 65 ENTER 2 OFF...
Страница 132: ...128 4 MicroDL 4 4 ENTER 1 5 MicroDL 5 5...
Страница 133: ...129 3 ENTER 3 SPCS SPCS SPCS 30 Microsoft Windows 95 CD ROM SPCS CareFusion MicroDL SPCS MicroDL SPCS MicroDL...
Страница 134: ...130 MicroDL MicroDL ENTER ENTER SPCS MicroDL COM SPCS MicroDL EN60950 1992 1993 SPCS SPCS SPCS SPCS...
Страница 135: ...131 MicroDL 10 9 PP3 24 CareFusion B T P S...
Страница 136: ...132 MicroDL BTPS 30 3 6 36 SM2125 MicroDL FVC...
Страница 137: ...133 Sidex 15...
Страница 139: ...135 I50 R50 PIF MVV Ind FET CareFusion ATS 9 3 10 24 190 170 70 60 0 40 30 90 RH 20 70 10 90 RH...
Страница 142: ......
Страница 143: ......