Appendix C Alarm Troubleshooting
Table 4 Alarm Troubleshooting
Alarm
Priority
Possible Cause
Actions
%O
2
< 18%
High
•
O
2
calibration
required.
•
Restore FiO
2
level to above the
minimum limit
•
Press Alarm Reset for 3 seconds.
•
Recalibrate O
2
as soon as
practicable.
%O
2
> 104% High
•
O
2
calibration
required.
•
Restore FiO
2
level to below the
maximum limit
•
Press Alarm Reset for 3 seconds.
•
Recalibrate O
2
as soon as
practicable.
High %O
2
(> 5% above
setting for 15
seconds).
High
•
Blender setting
changed.
•
Supply gas failure
•
Water trap overflow
•
Press Alarm Mute to silence the
alarm
•
Correct delivered oxygen
concentration
•
Press Alarm Reset for 3 seconds to
set new limits
Low %O
2
(> 5 % below
setting for 15
seconds).
High
•
Blender setting
changed.
•
Supply gas failure
•
Water trap overflow
•
Press Alarm Mute to silence the
alarm
•
Correct delivered oxygen
concentration
•
Press Alarm Reset for 3 seconds to
set new limits
Over
pressure
(> 11 cmH
2
O
in NCPAP
and BiPhasic
modes)
High
•
Flow rate set too
high.
•
Occlusion of
exhalation limb
•
Blocked
silencer/bacteria
filter
•
Check exhaust tube / filter
•
Reduce flow rate to achieve
pressure below high pressure limit
•
Press Alarm reset for 3 seconds to
set new limits
Over
pressure
(> 15 cmH
2
O
in BiPhasic
tr* mode)
High
•
Flow rate set too
high.
•
Occlusion of
exhalation limb
•
Blocked
silencer/bacteria
filter
•
Check exhaust tube / filter
•
Reduce flow rate to achieve
pressure below high pressure limit
•
Press Alarm reset for 3 seconds to
set new limits
Low battery
charge
(< 40%).
Medium/High
•
Battery status
indicator changes
from green to red.
•
Connect external power
Содержание Infant Flow SiPAP
Страница 1: ...Infant Flow SiPAP Operator s Manual ...
Страница 10: ...x Infant Flow SiPAP 675 101 101 Rev P Electromagnetic Environment Specifications ...
Страница 11: ...Infant Flow SiPAP xi 675 101 101 Rev P ...
Страница 14: ...xiv Infant Flow SiPAP 675 101 101 Rev P Infant Flow SiPAP ...
Страница 20: ...6 Chapter 2 Product Specifications 675 101 101 Rev P Infant Flow SiPAP ...
Страница 38: ...24 Chapter 5 Operation 675 101 101 Rev P Rear Panel Figure 8 Rear Panel ...
Страница 48: ...34 Chapter 5 Operation 675 101 101 Rev P Infant Flow SiPAP ...
Страница 64: ...50 Appendix A Product Configurations 675 101 101 Rev P Infant Flow SiPAP ...
Страница 65: ...Infant Flow SiPAP 51 675 101 101 Rev P Appendix B Pneumatic Diagram ...
Страница 66: ...52 Appendix B Pneumatic Diagram 675 101 101 Rev P Infant Flow SiPAP ...
Страница 70: ...56 Appendix C Alarm Troubleshooting 675 101 101 Rev P Infant Flow SiPAP ...
Страница 76: ...62 Appendix D Fault Management 675 101 101 Rev P Infant Flow SiPAP ...
Страница 78: ...64 675 101 101 Rev P Infant Flow SiPAP ...