Chapter 4 Unpacking and Setup
21
675–101–101 Rev. P
Perform the Alarms Test on the Infant Flow SiPAP driver using the following steps
and the initial settings provided above.
1. Make appropriate connections for air and oxygen gas supply. Connect power
cord to appropriate AC outlet. Attach patient circuit, generator and patient
interface (mask or prong) as shown in Figure 3. Occlude the opening to the
patient.
2. Power up the driver and allow Power On Check to complete.
3. Low airway pressure alarm: From NCPAP operating mode, with alarms set,
remove occlusion from opening to patient. Verify that the low airway pressure
alarm activates. Restore the patient interface occlusion and press the Alarm
Mute / Silence button for 3 seconds to reset the alarms.
4. High airway pressure alarm: Adjust the NCPAP / Pres Low flow meter to 11
L/min. Verify that the high airway pressure alarm activates. Return the NCPAP /
Pres Low flow meter to 8 L/min and press the Alarm Mute / Silence button for 3
seconds to reset the alarms.
5. High %O
2
Alarm: Adjust the % O
2
control to 35%. Verify that the High %O
2
alarm
activates. Return the O
2
control setting to 30%. Reset alarms by pressing the
Alarm Mute / Reset button for 3 seconds.
6. Low % O
2
Alarm: Adjust the % O
2
control to 25%. Verify that the Low %O
2
alarm
activates. Return the O
2
control setting to 30%. Reset alarms by pressing the
Alarm Mute / Reset button for 3 seconds.
7. Loss AC Alarm: Disconnect the AC power cord from the wall outlet. Verify that
the Loss AC alarm activates. Reconnect the AC power cord. Clear the alarm by
pressing the Alarm Mute / Reset button.
8. High Circuit Pressure Alarm: Increase nCPAP pressure to 11.1 cmH
2
O by
increasing the NCPAP/PRES Low Flow meter. Verify that the High Circuit
pressure alarm activates. Return NCPAP/PRES Low Flow meter to 8 LPM and
press the Alarm Mute/Silence button for three seconds to reset the alarms.
9. Low Breath Rate (Apnea) Alarm: Select and confirm BApnea/LBR (U.S.
configuration). Using the abdominal sensor, manually tap the abdominal sensor
to simulate a spontaneous breath rate. The default mandatory breath rate
should be left alone. No alarms should be present. Change the mandatory rate
control setting to 1 bpm and stop tapping the abdominal sensor. Verify that the
Low Breath rate alarm activates after the default interval of 20 seconds. Resume
simulating spontaneous breath rate, turn the rate control to the default setting
and clear the alarm by pressing the Alarm Mute/Reset button for 3 seconds.
Note: a transducer must be connected to perform the alarm check.
Содержание Infant Flow SiPAP
Страница 1: ...Infant Flow SiPAP Operator s Manual ...
Страница 10: ...x Infant Flow SiPAP 675 101 101 Rev P Electromagnetic Environment Specifications ...
Страница 11: ...Infant Flow SiPAP xi 675 101 101 Rev P ...
Страница 14: ...xiv Infant Flow SiPAP 675 101 101 Rev P Infant Flow SiPAP ...
Страница 20: ...6 Chapter 2 Product Specifications 675 101 101 Rev P Infant Flow SiPAP ...
Страница 38: ...24 Chapter 5 Operation 675 101 101 Rev P Rear Panel Figure 8 Rear Panel ...
Страница 48: ...34 Chapter 5 Operation 675 101 101 Rev P Infant Flow SiPAP ...
Страница 64: ...50 Appendix A Product Configurations 675 101 101 Rev P Infant Flow SiPAP ...
Страница 65: ...Infant Flow SiPAP 51 675 101 101 Rev P Appendix B Pneumatic Diagram ...
Страница 66: ...52 Appendix B Pneumatic Diagram 675 101 101 Rev P Infant Flow SiPAP ...
Страница 70: ...56 Appendix C Alarm Troubleshooting 675 101 101 Rev P Infant Flow SiPAP ...
Страница 76: ...62 Appendix D Fault Management 675 101 101 Rev P Infant Flow SiPAP ...
Страница 78: ...64 675 101 101 Rev P Infant Flow SiPAP ...