27
Safety Designation per IEC 60601-1
Type of protection against
electrical shock
Internally powered
Equipment
Degree of protection against
electrical shock
Type B applied part
Power Equipment
Battery type: PP3 battery,
Alkaline Manganese
Dioxide, 9.0V, 550mAh
Battery life:
2000 tests.
Degree of Electrical connection
between equipment and Patient
Equipment designed as
non-electrical connection to
the patient.
Degree of mobility
Transportable
Mode of operation
Continuous
Classification according to IEC 60601-1
BabyCO Meter
Applied part, type B
WARNING:
No modification of this equipment is allowed.
WARNING:
Do not connect devices that are not specified as part of
the system.
NOTE:
When you connect other equipment to the unit, always make
sure that the whole combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems. When
connecting to a PC with the supplied serial cable then the PC must
be IEC 60601-1 / ANSI/AAMI ES60601-1:2005 / CAN/CSA-C22.2
No. 60601-1:14 compliant.
WARNING:
The user must not touch any voltage-carrying parts and
the patient at the same time.
Содержание BabyCO Meter
Страница 2: ......
Страница 10: ...8 ...
Страница 41: ...39 ...
Страница 42: ...40 ...
Страница 47: ...45 6 5 4 1 2 3 CO FCOHb OFF ...
Страница 50: ...48 Das BabyCO ist zur Verwendung durch Ärzte und medizinisches Fachpersonal vorgesehen ...
Страница 79: ...77 ...
Страница 80: ...78 ...
Страница 114: ...112 ...
Страница 146: ...144 ...
Страница 181: ...179 ...
Страница 182: ...180 ...
Страница 216: ...214 ...
Страница 221: ...219 6 5 4 1 2 3 CO FCOHb OFF ...
Страница 254: ...252 ...
Страница 264: ...262 将滑动开关推至 CO PPM 位置 等待设备显示清零 如下所示 将控制阀紧紧拧到气缸上 并连接气源 串行 定标 开关 ...