CareFusion Alaris PK Plus MK4 8005PK201 Скачать руководство пользователя страница 3

1000DF00741 Issue 5

2/52

Alaris™ PK Plus Syringe Pump

Introduction

Introduction

This Directions for use can be used with the Alaris™ PK Plus Syringe Pump MK4.

w

The Pumps can be identified as MK4 version by the MK4 on 
the label on the rear case, see image right, or by verifying the 
software version as 3.3.0 or above on power up.

The Alaris PK Plus Syringe Pump (hereinafter referred to as Pump) provides the user with an infusion tool for the administration of drugs 
for anaesthesia. The embedded software within the Pump is loaded with three compartment pharmacokinetic predictive models and 
has 4 modes of operation:
1.   Continuous infusion (ml/h)
2.   Total Intravenous Anaesthesia (TIVA) mode.  

 

– In this mode the user is able to select the infusion rate and administer bolus doses as required.

3.  Total Intravenous Anaesthesia (TIVA) with TCI predictions mode. 

 

– In this mode the user is able to select the infusion rate and administer bolus doses as required. The pharmacokinetic model is 

used to estimate the plasma and effect site concentration.

4.  TCI Mode

•  Plasma target-controlled infusion (TCI).

 

– In this mode the user selects the desired (target) plasma drug concentration, and the pharmacokinetic model is used to 

calculate the infusion rates required to achieve that concentration. A graphic display shows the trajectory of the estimated 
plasma and effect site drug concentration over time.

•  Effect Site target-controlled infusion (TCI).

 

– In this mode the user sets the desired effect site target concentration and the pharmacodynamic model is used to calculate 

the infusion rates required to achieve that concentration. A graphic display shows the trajectory of the estimated effect site 
and plasma concentration over time.

The Alaris PK Plus Syringe Pump has a user friendly interface that displays the infusion rate, the total drug dose delivered, and the 
estimated plasma and effect-site concentrations to enable the user to follow the drug prescription information from the relevant 
country.

Intended Purpose

The Alaris PK Plus Syringe Pump is intended for use by medical staff for purposes of controlling infusion rate and volume.

Conditions of Use

The Alaris PK Plus Syringe Pump should only be operated by a clinician competent in use of automated syringe pumps and post-
placement management of intravenous catheters.  
Use of the Alaris PK Plus Syringe Pump does not limit the responsibility of the anaesthetist for drugs administration. It is important that 
users operating the Alaris PK Plus Syringe Pump are fully aware of the available literature for any model used in association with a drug 
and that they refer to the prescribed information for rate and dosing limits. Pharmacokinetic and Pharmacodynamic Interactions among 
anaesthetic drugs are known, but are not taken into account in the calculation of the plasma and effect site concentrations.
The user should be appropriately trained in the use of the Pump and should follow the recommendations of this Direction For Use 
(DFU).
In particular, the user must be aware that starting the Pump in a TCI mode will result in the automatic infusion of a pre-calculated bolus 
dose followed by an infusion to achieve the selected target concentration. The initial parameter calculations are displayed on screen 
prior to starting the infusion. It is thus essential that the user verifies that the patient characteristics and the selected infusion rate or 
target concentration conform with the drug prescribing information of the relevant country.
CareFusion has verified the accuracy of the mathematical model implementation as well as pump delivery accuracy - (specification and 
accuracy of pump - delivery are available in 'Profiles from TCI Mode' section).
Different drugs are associated with dedicated models – each model consists of a set of standard pharmacokinetic parameters which can 
be selected and used by the embedded 3 compartment model used in the Alaris PK Plus Syringe Pump (where use of that drug in TCI 
mode is authorised);
Diprivan from ASTRA-ZENECA is the only recommended Propofol formulation to be used in TCI mode as per prescribing information. 
This Pump includes the Marsh model for the calculation of the Diprivan infusion rates, and plasma and effect-site concentrations.
When Remifentanil and Sufentanil are used in TCI mode, – the Minto and Gepts models respectively – are used to calculate the required 
infusion rates.

Содержание Alaris PK Plus MK4 8005PK201

Страница 1: ...s Alaris PK Plus Syringe Pump MK4 Model 8005PK201 Directions For Use en...

Страница 2: ...s 13 Getting Started 16 Syringe Loading 18 Starting the Pump 21 Basic Features 23 Operations During Use 25 Alarms and Warnings 27 Prompts 28 Configured Options 29 Specifications 33 Recognised Syringes...

Страница 3: ...rolling infusion rate and volume Conditions of Use The Alaris PK Plus Syringe Pump should only be operated by a clinician competent in use of automated syringe pumps and post placement management of i...

Страница 4: ...Where stated a minimum infusion rate refers to a nominal rate of 1 0ml h and an intermediate infusion rate refers to a nominal rate of 5 0ml h The complete range of infusion rates settings and values...

Страница 5: ...rate constants k12 k21 k31 and k13 or clearances Drug metabolism is described by the rate constant k10 Fig 2 The aim of TCI techniques is to use pharmacokinetic modelling to calculate the infusion rat...

Страница 6: ...able titration to effect if necessary Note Specific model parameters are available in the TCI Overview section or directly on the Pump via the information key when selecting drugs Users should refer t...

Страница 7: ...n 20 ng ml Vc 5 1 0 0201 x age 40 0 072 x lbm 55 V2 9 82 0 0811 x age 40 0 108 x lbm 55 V3 5 42 cl1 2 6 0 0162 x age 40 0 0191 x lbm 55 cl2 2 05 0 0301 x age 40 cl3 0 076 0 00113 x age 40 k10 cl1 Vc k...

Страница 8: ...d as proof of approval by an authorised person according to Hospital protocol Signed printout to be kept safe for use during verification procedure Release Data Set status to be promoted to Released p...

Страница 9: ...ay Syringe Clamp ON OFF RUN HOLD PURGE BOLUS MUTE PRESSURE OPTION Extension set hook Finger Grips Extension set hook Release lever for Rotating Cam Carrying Handle IR Communications port RS232 Connect...

Страница 10: ...PURGE BOLUS button Press to access PURGE or BOLUS softkeys Press and hold down softkey to operate PURGE the extension set during set up Pump is on hold Extension set must not be connected to the pati...

Страница 11: ...gainst direct sprays of water up to 15 from vertical and protected against solid objects greater than 2 5mm Note IP33 applies if AC power cable retainer kit part number 1000SP01294 is fitted r Alterna...

Страница 12: ...Rate Initial Maintenance Rate Pause Before Maintenance Induction Duration Time of Induction Initial Induction Dose Initial Induction Volume TCI Mode MORE Information Screen Selecting the MORE softkey...

Страница 13: ...n screen D Duration of Induction Phase Displayed on protocol confirmation screen E Duration of Hands Free Bolus Displayed in bolus set up screen F Maintenance Phase Dose Rate Displayed on protocol con...

Страница 14: ...otect against the introduction of air the user should regularly monitor the progress of the infusion syringe extension line and patient connections and follow the priming procedure specified herein Op...

Страница 15: ...amaged remove it from service for inspection by Qualified Service Personnel When transporting or storing the Pump use original packaging where possible and adhere to temperature humidity and pressure...

Страница 16: ...s recommendations regarding electromagnetic interference EMI For further information please refer to the product Technical Service Manual TSM Alternatively contact your local CareFusion representativ...

Страница 17: ...3 Connect the Pump to the AC Power Supply for at least 2 hours to ensure that the internal battery is charged verify that the S is lit Language Selection 1 On initial start up the Pump will display th...

Страница 18: ...ively loose movement in the extended mountable position If these signs are observed the Pumps should be taken out of service for examination by Qualified Service Personnel Docking Station Workstation...

Страница 19: ...not connected to the patient Practice Recommendations Tubing internal diameter Smallbore or microbore tubing is recommended when infusing at low rates Filters Internal volume dead space of in line fil...

Страница 20: ...in the extension set and syringe at the end of infusion as this cannot be fully infused Syringe Flange Clamp Syringe Clamp Barrel Flange Plunger Grippers Plunger Flange Plunger Holder Plunger Finger G...

Страница 21: ...can be changed by pressing the TYPE softkey CONFIRM TYPE IVAC 50 ON HOLD Note If the PURGE SYRINGE option has been enabled then the prompt to purge screen is displayed and the extension set can be pu...

Страница 22: ...for the selected drug must be entered using the f keys and press the OK softkey to confirm The required parameters may include the following depending on the model AGE HEIGHT GENDER LBM and BMI Lean B...

Страница 23: ...ion time The cap rate will be cleared when first titration occurs 11 Target Concentration Cpt or Cet Adjust the Target Concentration if necessary using the f keys Confirm the Target Concentration and...

Страница 24: ...volume has been delivered or the bolus volume limit is reached release the softkey The bolus volume is added to the total volume infused BOLUS Enabled Hands Free The Hands Free Bolus is delivered wit...

Страница 25: ...Rate Titration is enabled the rate can be adjusted while infusing 1 Select the new rate using the f keys 2 The message START TO CONFIRM will display on screen titration callback tone will sound and P...

Страница 26: ...predictive TIVA mode the initial doserate will be set to zero DECREMENT CONC In TCI mode only 1 Press the d button to access the options menu 2 Select DECREMENT CONC 3 Select the required DECREMENT C...

Страница 27: ...rate To precisely set a doserate the arrow must be pointing to the doserate mg kg h the flowrate will be calculated from the doserate To precisely set a flowrate the arrow must be pointing to the flo...

Страница 28: ...30 minutes a continuous audible alarm will indicate that the battery is exhausted Reconnect to the AC Power Supply to continue operation and charge the internal battery BATTERY EMPTY High Red The int...

Страница 29: ...ing the YES softkey If OVERRIDE LIMIT is not required press the NO softkey and adjust rate below Soft Alert BOLUS DOSE OVER The bolus dose has been set to a value which exceeds a Soft Alert Check the...

Страница 30: ...CONTRAST from the Configured Options menu using the f keys and press the OK softkey 2 Use the f keys to select a contrast ratio value The contrast of the display will change when scrolling through th...

Страница 31: ...so need to be confirmed Pressure Default The default occlusion pressure alarm level Pressure Display Sets whether the Pressure Information is available on the main display Purge Rate The rate used dur...

Страница 32: ...ent target concentration TIVA Induction Parameters Induction ON OFF Enables Disables induction stage of TIVA protocol Dosing Units The induction dose units This can be based on patient weight Default...

Страница 33: ...g kg minutes 1 g kg h Maintenance Hard Max 20mg kg h 20mg kg h 2 g kg minutes 2 g kg minutes 2 g kg h Default Bolus Rate 1200ml h 600ml h 600ml h 600ml h 1200ml h Default Bolus 1 0mg kg 1 0mg kg 1 0 g...

Страница 34: ...aximum 100 0ml Increments of 0 1ml default 5 0ml During BOLUS the pressure limit alarms are temporarily increased to their maximum level Bolus Volume Accuracy Selected maximum rates are shown below Bo...

Страница 35: ...r force in recognised syringes can cause variations in accuracy and trumpet curves See also trumpet curves section in this manual Electrical Classification Class I product Continuous Mode Operation Tr...

Страница 36: ...ng Relative Humidity 20 90 Operating Atmospheric Pressure 700hPa 1060hPa Transport and Storage Temperature 30 C 50 C Transport and Storage Relative Humidity 10 95 Transport and Storage Atmospheric Pre...

Страница 37: ...ing and Confirming a Syringe section 2 TYCO Healthcare KENDALL MONOJECT w To minimise the risk of incorrect confirmation of the syringe type it is recommended that only syringe types available in the...

Страница 38: ...aximum Protection Against Electrical Shock Class 1 Classification Continuous Operation Supply to Pump 115 230V 50 60Hz 60VA The Alaris DS Docking Station Product SKU 80283UNS00 xx Supply Voltage 230VA...

Страница 39: ...priming volume 13cm 20062E7D 3 way extension set with 3 SmartSite Needle Free Valves and one backcheck valve 16cm MFX2271 2 way set with anti syphon valve and backcheck valve 210cm MFX2270 3 way set w...

Страница 40: ...ector set 3 way with 2 anti syphon valves and backcheck valve 190cm 500 002V Y Connector set 2 way with anti syphon valve and backcheck valve 15cm 500 003V Y Connector set 3 way with 2 anti syphon val...

Страница 41: ...Standard PVC Syringe Extension Set 150cm Priming Volume 2 6ml G40020B Standard PVC Syringe Extension Set 200cm Priming Volume 1 4ml G40320V Opaque White PVC Syringe Extension Set 200cm Priming Volume...

Страница 42: ...used in the calibration procedure are standard SI The International System of Units units Battery Operation The internal rechargeable battery allows continued operation when the AC power is unavailab...

Страница 43: ...the Pump has visible cracks or damage to the case do not clean and immediately take it out of service for examination by Qualified Service Personnel The syringe and extension sets are disposable sing...

Страница 44: ...expected in the event of an occlusion when the BD Plastipak 50 ml syringe is selected with a G40020B standard extension set Time to alarm 1 0 ml h Time to alarm 5 0 ml h hr min typical hr min typical...

Страница 45: ...e to control the syringe Pump using the RS232 interface at some distance from the Pump and hence remote from the patient responsibility for the control of the Pump is vested in the software run on the...

Страница 46: ...nimum Enable Active Low 7V to 12V powers up the isolated RS232 circuitry Active High 7V to 12V Inactive Floating open circuit allows isolated RS232 circuitry to power down Isolation Socket Pump 1 5kV...

Страница 47: ...reduced short term fluctuations have greater effects as represented by the mouth of the trumpet Knowledge of system accuracy over various observation windows may be of interest when certain drugs are...

Страница 48: ...mean volumetric accuracy in TCI Mode better than 5 2 By measuring the volume from the flow rate profile delivered from the Alaris PK Plus Syringe Pump and then introducing this into a reverse pharmac...

Страница 49: ...ge Remifentanil Minto Model BD 50ml Syringe Patient Age 75 Years Patient Weight 65kg Patient Height 175cm Patient Gender Male Drug Concentration 50 g ml Volumetric Accuracy 0 2 Patient Age 75 Years Pa...

Страница 50: ...tanil Minto Model BD 50ml Syringe Patient Age 75 Years Patient Weight 65kg Patient Height 175cm Patient Gender Male Drug Concentration 50 g ml Plasma Concentration Accuracy 0 2 Patient Age 75 Years Pa...

Страница 51: ...he World Wide Web at www carefusion co uk alaris technical A username and password are required to access our manuals Please contact a local customer services representative to obtain login details Pa...

Страница 52: ...8 10 3994 DB Houten Nederland CareFusion 10020 Pacific Mesa Blvd San Diego CA 92121 USA Tel 32 0 2 267 38 99 Tel 34 902 555 660 Tel 31 0 30 2289 711 Tel 1 800 854 7128 Fax 32 0 2 267 99 21 Fax 34 902...

Страница 53: ...1000DF00741 Issue 5 52 52 Alaris PK Plus Syringe Pump Service Contacts Page Intentionally Left Blank...

Страница 54: ...n of CareFusion Corporation or one of its affiliates and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sell any product described Reproduction dis...

Отзывы: