1000DF00741 Issue 5
26/52
Alaris™ PK Plus Syringe Pump
Operations During Use
24 Hour Log
This option allows the 24 hour log of volume infused to be reviewed.
1. Press the
d
button to access the options menu.
2. Select the
24H LOG
option using the
f
keys and press the
OK
softkey.
The display shows the hourly volume infused. The volume infused shown in brackets is the total volume infused since the volume was
last cleared. See example below:
07:48 - 08:00 4.34ml (4.34ml)
08:00 - 09:00 2.10ml (6.44ml)
09:00 - 10:00 2.10ml (8.54ml)
VOLUME CLEARED
3. Press the
QUIT
softkey to exit the log.
Event Log
This option allows the event log to be reviewed. It can be enabled/disabled.
1. Press the
d
button to access the options menu.
2. Select the
EVENT LOG
option using the
f
keys and press the
OK
softkey.
3. Scroll through the log using the
f
keys. Press the
QUIT
softkey to exit the log.
Data Set Details
To review currently selected data set information:
1. Press the
d
button to access the options menu.
2. Select
DATA SET DETAILS
.
3. Review the information and press the
QUIT
softkey to exit.
SET BY DOSERATE/SET BY ml/h (TIVA mode only)
To set doserate to flowrate in precise increments, it may be necessary to switch between the rate adjust options
SET BY DOSERATE
and
SET BY ml/h
. An arrow to the left of the rate display shows the rate changed when the
f
keys are used to increase/decrease
the infusion rate.
To precisely set a doserate, the arrow must be pointing to the doserate (mg/kg/h); the flowrate will be calculated from the doserate. To
precisely set a flowrate, the arrow must be pointing to the flowrate (ml/h); the doserate will be calculated from the flowrate.
Selecting the
SET BY ml/h
option:
1. Whilst the Pump is infusing, press the
d
button to access the options menu.
2. Select the
SET BY ml/h
option using the
f
keys and press the
OK
softkey indicated on the screen. This will select the
SET BY FLOWRATE
option, the arrow on the display will automatically select the flowrate, the flowrate can be adjusted if required.
Selecting the
SET BY DOSERATE
option:
1. Whilst the Pump is infusing, press the
d
button to access the options menu.
2. Select the
SET BY DOSERATE
option using the
f
keys and press the
OK
softkey indicated on the screen. This will select the
SET BY DOSERATE
option, the arrow on the display will automatically select the doserate, the doserate can be adjusted if required.
EFFECT SITE TCI
When in
PLASMA TCI
mode the user is able to switch to
EFFECT SITE TCI
mode if the configuration permits:
1. Press the
d
button to access the options menu.
2. Select
EFFECT SITE TCI
using the
f
keys.
3. Press the
OK
softkey indicated on the screen. A confirmation screen will be displayed.
PLASMA TCI
When in
EFFECT SITE TCI
mode the user is able to switch to
PLASMA TCI
mode if the configuration permits:
1. Press the
d
button to access the options menu.
2. Select
PLASMA TCI
using the
f
keys.
3. Press the
OK
softkey indicated on the screen. A confirmation screen will be displayed.