5-15
Alaris
®
SpO
2
Modules
Technical Service Manual
CORRECTIVE MAINTENANCE
Table 5-3. Level of Testing Guidelines
Use the Maintenance Software to perform testing.
Tests to Perform
l
= Required
Blank = Not Applicable
Repair/Replacement of:
Alarm T
est
Channel ID / IUI Connector
T
est
Display T
est
Keypad T
est
Patient Cable
Alarm
T
est
Patient Lead Electrical Leakage
Power On Self
T
est
Pulse Rate / Saturation
T
est
Speaker T
est
V
isual / Instrument Inspection
Display Board
l
l
l
l
l
l
l
l
Front Case / Keypad
l
l
Logic Board
l
l
l
l
l
l
l
l
l
SpO
2
Board
l
l
l
Patient Connector Board
l
l
Power Supply Board
l
l
l
l
l
l
l
Rear Case / IUI Connector
l
l
Miscellaneous:
Instrument Dropped
l
l
l
l
l
l
l
l
l
l
New Instrument Checkout
l
l
l
l
l
No Fault Found
(
instrument not opened
)
l
l
l
l
l
l
l
l
l
l
No Fault Found
(
instrument opened
)
l
l
l
l
l
l
l
l
l
l
Содержание Alaris 8210 Series
Страница 12: ...2 6 Alaris SpO2 Modules Technical Service Manual CHECKOUT AND CONFIGURATION THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 14: ...3 2 Alaris SpO2 Modules Technical Service Manual PREVENTIVE MAINTENANCE THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 32: ...5 16 Alaris SpO2 Modules Technical Service Manual CORRECTIVE MAINTENANCE THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 36: ...6 4 Alaris SpO2 Modules Technical Service Manual TROUBLESHOOTING THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 55: ......