 is an et medical devices SpA brand 

w w w . c a r d i o l i n e . b i z

 

4 Testing the safety characteristics 

The safety regulations envisage two important tests: 

  The leakage currents test measures the value of the currents lost in relation to the safety of the patient and the operator. 

Warning: All safety tests must be performed according to standards EN.60601-1(1990 paragraphs 19 - 20) 

EN 60601-2-25 (1995) unless otherwise specified in the local safety regulations.  

4.1  Leakage currents test 

Warning:  This  test  must  be  performed  every  time  the  device  has  been  opened  for  inspection  and/or 

repair, and in any event every two years, unless otherwise specified by the local safety regulations.   

Connect the electrocardiogram to the battery charger, and then connect this assembly to the measuring 

instrument according to the instrument’s manual, recalling that: 

'

 

The  leakage  current  to  the  casing  is  measuring  between  the  mains  supply  circuits  and  a  metal  sheet  no  greater  than  20  x  10  cm 

pressed against the casing of the device.  

'

 

The leakage current in the patient is measured between the mains and the applied part. For connection to the applied part use the 

patient lead itself. 

"'

 

The leakage current in the patient with mains voltage directly on the applied part (first failure condition) is measured between the meta 

sheet connected to the device and the applied part. 

'

 

The auxiliary current in the patient is measured singly on each electrode (excluding the reference electrode) compared to all the other 

electrodes connected together. 

 

Note: Make the measurements following the indications in the instrument user manual, and check that the leakage current values 

measured are less than or equal to those listed in table IV. 

Table IV. 
Admissible permanent values for leakage and auxiliary currents in the patient in mA (milliamperes). 

 

: )

&3)

, -

2

, -

2

2

, -

<

2 2

2 2

, -
%

'

77777

2 2

9 =

!

<

2 2

2 2

: )

> :

&3) > 3

 

Содержание 80509531

Страница 1: ...w w c a r d i o l i n e b i z T TE EC CH HN NI IC CA AL L S SE ER RV VI IC CE E M MA AN NU UA AL L E EL LE EC CT TR RO OC CA AR RD DI IO OG GR RA AP PH H M MO OD DE EL L C Ca ar rd di io ol li in ne...

Страница 2: ...2 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z Index 0 1 2 3 4 2 3 0 1 4 5 1 6...

Страница 3: ...3 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 2 1 7 2 2 2 3 2 2 2 2 2 4 8 9 0 9 0 9 0 3...

Страница 4: ...o store up to 40 full ECG exams with no need to print out immediately on paper paper saving mode ECG measurements option automatic ECG parameter measurement program ECG signal interpretive option a us...

Страница 5: ...that the signal is restored use only original electrodes or electrodes responding to IEC and AAMI standards If an electrosurgical scalpel is in use the patient cable should be disconnected from the d...

Страница 6: ...MH 12V 2 3 Display Display assembly is composed of frame monochromatic LCD display 320 x 240 pixel 4 7 inch as well as a step up converter 2 4 Keyboard The keyboard card consists of the device functio...

Страница 7: ...to the patient input port 8 4 1 2 3 6 7 5 15 9 10 11 12 13 14 Port from connection side Pin 1 IN C2 electrode C2 Pin 2 IN C3 electrode C3 Pin 3 IN C4 electrode C4 Pin 4 IN C5 electrode C5 Pin 5 IN C6...

Страница 8: ...nstrument s manual recalling that The leakage current to the casing is measuring between the mains supply circuits and a metal sheet no greater than 20 x 10 cm pressed against the casing of the device...

Страница 9: ...nnected to the device to the positive pin of instrument 5 1 a connect all the other cable terminals to the negative pin of instrument 5 1 a record the signal for a few seconds check that the amplitude...

Страница 10: ...the signal on leads V1 V6 measure the length of the wave cycle recorded on the paper The results should be as follows Period 50 mm 5 for a feed rate of 50 mm s Period 25 mm 5 for a feed rate of 25 mm...

Страница 11: ...3 2 0 0 03 4 2 03 4 3 2 2 2C 3 2 22 2C 7 2C 28 3 22 2 2C 7 2C Note The 0 5 Hz pitch linear phase anti drift filter is always on and cannot be switched off The 50 or 60 Hz filter eliminates modified no...

Страница 12: ...illustrated in the picture below Changing mains voltage does not require mains fuses rate modification 5 Warning Changing mains voltage imply the replacement of the mentioned label Only original ones...

Страница 13: ...nal polarisation circuit Hybrid circuit for amplification filter and clamp 6 1 6 Motor control This consists of the following parts Stroboscopic sensor control Phase comparator between reference frequ...

Страница 14: ...d 6 4 Keyboard card The keyboard consists of 18 function keys 10 alphanumeric keys 1 LED for mains line indication CPU connection through bi directional synchronous serial port Testing the keyboard ca...

Страница 15: ...15 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z...

Страница 16: ...mains fuses if OK replace motherboard Check efficiency of battery and replace if necessary Check T5A fuse on motherboard Keyboard keys not working The key pressed not enabled for the specific function...

Страница 17: ...nd display assembly Verify on display connector J49 presence of LCD data signal level from 0 trough 5 V if present replace display assembly otherwise replace mother board 8 How to dismantle and reasse...

Страница 18: ...sembly proceed in inverse order 8 8 Removing the printer assembly table T2b Proceed as follows proceed as in chapter 8 3 for opening the equipment and removing the mother board from the equipment remo...

Страница 19: ...pull it off the casing If any adhesive is left on the casing you must remove it by rubbing with your fingertips To fit a new plate center it with the corners in the space provided and press gently ov...

Страница 20: ...s is the numerical position of the digital potentiometer 9 1 3 Speed calibration Testing and setting Select and firm the SPEED CAL 2 submenus and the speed to calibrate The following messages then app...

Страница 21: ...requency of controls required based on normal use of the electrocardiograph about 4000 ECG recordings per year Warning Check immediately if there are events not referring to normal use General and saf...

Страница 22: ...thermal head correctly use a cloth slightly dampened with alcohol or an equivalent solution with the device off Proceed as follows open the paper feed pass the cloth over the dots of the thermal head...

Страница 23: ...R 1200 VIEW 69701441 DISPLAY GLASS FOR AR 1200 VIEW 69701492 ELECTRONIC KEYBOARD BOARD AR 1200 VIEW EXT 69701502 MOTHER BOARD IRDA AR1200VIEW EXT C LINE 69701502E MOTHER BOARD IRDA AR1200VIEW EXT C LI...

Страница 24: ...s the insulation required for connection with medical systems The communication with a PC occurs through a proprietary protocol a standard IRDA protocol cannot be used et medical devices has validated...

Страница 25: ...ime constant 3 3 s Acquisition 11 bit 1000 samples s channel printing and filters 500 samples s channel in calculation and filters Resolution 5 V bit Leads 9 9 9 0 9 8 acquired 4 derived III aVR aVL a...

Страница 26: ...g in real time Automatic simultaneous acquisition ECG Autotimer acquisition at user defined intervals Paper Saving signal acquisition in simultaneous mode with possibility of copying or storing on PC...

Страница 27: ...storage Environment temperature from 10 C to 40 C Relative humidity from 10 to 95 without condensation Atmospheric pressure from 500 to 1060 hPa Dimensions 320 x 72 x 240 mm length x height x depth W...

Страница 28: ...onductive material and earthed The person performing the repair must be earthed with a suitable protective bracelet compliant with safety regulations The material must be contained in suitable antista...

Страница 29: ...29 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 16 APPENDIX B 16 1 Figures and illustrated tables 0 D...

Страница 30: ...DE VIEW VISTA LATERALE SIDE VIEW INGRESSO PAZIENTE PATIENT INPUT TASTO RESET RESET KEY 1 1 2 3 4 BATTE RY PACK 1 2 VITI FISSAGGIO FONDO MOBILE LOWER HOUSING FIXING SCREWS VITE FISSAGGIO SPORTELLO VANO...

Страница 31: ...ORE TESTA TERMICA THERMAL HEAD FLAT CONNECTOR FLAT CONNETTORE DISPLAY DISPLAY FLAT CONNECTOR 6 8 7 9 FLAT CONNETTORE TASTIERA KEYBOARD FLAT CONNECTOR INTERFACCIA SERIALE I R SERIAL INTERFACE I R FLAT...

Страница 32: ...I FISSAGGIO SCHEDA MADRE MOTHER BOARD FIXING SCREWS 2 1 3 4 5 VITI FISSAGGIO SCHEDA TASTIERA KEYBOARD FIXING SCREWS SCHEDA TAST IERA KEYBOARD VITI FISSAGGIO SISTEMA SCRIVENTE WRITING SYSTEM FIXING SCR...

Страница 33: ...COLARI DETAILS VIEW CHIMOGRAFO MOTOR ASSEMBLY T3 FLAT STAMPANTE FLAT THERMAL HEAD SPORTELLO CARTA PAPER FEED SISTEMA SCRIVENTE WRITING SYSTEM RULLO GUIDA CARTA PAPER GUIDE ROLLER MOZZO PORTA CARTA PAP...

Страница 34: ...a testa termica ruotando in senso orario o antiorario l apposito ingranaggio max 1 giro dell attrezzo Nel solo caso di stampa non uniforme procedere come segue WRITER MOUNTING INSTRUCTIONS Note Remove...

Страница 35: ...N 22 e 23 di J49 STROBE 1 e 2 STROBE 3 e 4 PIN 7 DI J49 LATCH tsetup LATCH thold LATCH thold DATA tsetup DATA 333 nSEC DATA PIN 6 DI J49 CLOCK 3MHz 1 msec TEST POINT TESTA TERMICA THERMAL HEAD TEST PO...

Страница 36: ...DIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 17 Revision Sheet Revision Sheet Service Manual ECG AR 1200 VIEW Lang English CODE 80509531 80509532 REV OM DESCRIPTION DATE...

Страница 37: ...an et medical devices SpA brand w w w c a r d i o l i n e b i z DIAGRAMS LIST ECG ar 1200 view CARDIOLINE 115 V 115 V 50 60 Hz Code 80509531 80509532 3 0 1 0 0 1 0 2 222 77777777777777 222 222 222 222...

Страница 38: ...38 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z...

Отзывы: