11   Cleaning and disinfection 

11.1  Cleaning the device, electrodes and patient cable 

 

Device:

 use a cloth dampened with water or ethyl alcohol. Do not use other chemical products or household detergents. 

 

Electrodes:

 remove the electrodes from the patient cable and wash under running water. Do not scratch the electrodes and do 

not wash the leads box and the patient socket. 

 

Patient cable:

 do not immerse in water, use a cloth dampened with alcohol or an equivalent solvent. 

Note: the device cannot be sterilized! The electrodes can be sterilized with ethylene oxide. 

11.2  Clean thermal head 

To clean the printer thermal head correctly use a cloth slightly dampened with alcohol or an equivalent solution 

with the device off. 
Proceed as follows: 

  open the paper feed; 

  pass the cloth over the dots of the thermal head without pressing too hard. 

Warning: the printer thermal head is extremely sensitive to electrostatic potentials. Do not touch it for any 

reason. If necessary, handle it after connecting to a ground via a suitable protective bracelet or belt. 

11.3  Cleaning the paper feed roller 

To clean the paper feed roller correctly, remove it and use a soft cloth lightly dampened with alcohol or an 

equivalent solvent, turning the roller to clean the whole surface. 

11.4  Cleaning the mark sensor 

To clean the mark sensor correctly use a cloth slightly dampened with alcohol or an equivalent solution with the 

device off. 
Proceed as follows: 

  remove the paper feed; 

  pass the cloth over mark sensor without pressing. 

Содержание 80509531

Страница 1: ...w w c a r d i o l i n e b i z T TE EC CH HN NI IC CA AL L S SE ER RV VI IC CE E M MA AN NU UA AL L E EL LE EC CT TR RO OC CA AR RD DI IO OG GR RA AP PH H M MO OD DE EL L C Ca ar rd di io ol li in ne...

Страница 2: ...2 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z Index 0 1 2 3 4 2 3 0 1 4 5 1 6...

Страница 3: ...3 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 2 1 7 2 2 2 3 2 2 2 2 2 4 8 9 0 9 0 9 0 3...

Страница 4: ...o store up to 40 full ECG exams with no need to print out immediately on paper paper saving mode ECG measurements option automatic ECG parameter measurement program ECG signal interpretive option a us...

Страница 5: ...that the signal is restored use only original electrodes or electrodes responding to IEC and AAMI standards If an electrosurgical scalpel is in use the patient cable should be disconnected from the d...

Страница 6: ...MH 12V 2 3 Display Display assembly is composed of frame monochromatic LCD display 320 x 240 pixel 4 7 inch as well as a step up converter 2 4 Keyboard The keyboard card consists of the device functio...

Страница 7: ...to the patient input port 8 4 1 2 3 6 7 5 15 9 10 11 12 13 14 Port from connection side Pin 1 IN C2 electrode C2 Pin 2 IN C3 electrode C3 Pin 3 IN C4 electrode C4 Pin 4 IN C5 electrode C5 Pin 5 IN C6...

Страница 8: ...nstrument s manual recalling that The leakage current to the casing is measuring between the mains supply circuits and a metal sheet no greater than 20 x 10 cm pressed against the casing of the device...

Страница 9: ...nnected to the device to the positive pin of instrument 5 1 a connect all the other cable terminals to the negative pin of instrument 5 1 a record the signal for a few seconds check that the amplitude...

Страница 10: ...the signal on leads V1 V6 measure the length of the wave cycle recorded on the paper The results should be as follows Period 50 mm 5 for a feed rate of 50 mm s Period 25 mm 5 for a feed rate of 25 mm...

Страница 11: ...3 2 0 0 03 4 2 03 4 3 2 2 2C 3 2 22 2C 7 2C 28 3 22 2 2C 7 2C Note The 0 5 Hz pitch linear phase anti drift filter is always on and cannot be switched off The 50 or 60 Hz filter eliminates modified no...

Страница 12: ...illustrated in the picture below Changing mains voltage does not require mains fuses rate modification 5 Warning Changing mains voltage imply the replacement of the mentioned label Only original ones...

Страница 13: ...nal polarisation circuit Hybrid circuit for amplification filter and clamp 6 1 6 Motor control This consists of the following parts Stroboscopic sensor control Phase comparator between reference frequ...

Страница 14: ...d 6 4 Keyboard card The keyboard consists of 18 function keys 10 alphanumeric keys 1 LED for mains line indication CPU connection through bi directional synchronous serial port Testing the keyboard ca...

Страница 15: ...15 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z...

Страница 16: ...mains fuses if OK replace motherboard Check efficiency of battery and replace if necessary Check T5A fuse on motherboard Keyboard keys not working The key pressed not enabled for the specific function...

Страница 17: ...nd display assembly Verify on display connector J49 presence of LCD data signal level from 0 trough 5 V if present replace display assembly otherwise replace mother board 8 How to dismantle and reasse...

Страница 18: ...sembly proceed in inverse order 8 8 Removing the printer assembly table T2b Proceed as follows proceed as in chapter 8 3 for opening the equipment and removing the mother board from the equipment remo...

Страница 19: ...pull it off the casing If any adhesive is left on the casing you must remove it by rubbing with your fingertips To fit a new plate center it with the corners in the space provided and press gently ov...

Страница 20: ...s is the numerical position of the digital potentiometer 9 1 3 Speed calibration Testing and setting Select and firm the SPEED CAL 2 submenus and the speed to calibrate The following messages then app...

Страница 21: ...requency of controls required based on normal use of the electrocardiograph about 4000 ECG recordings per year Warning Check immediately if there are events not referring to normal use General and saf...

Страница 22: ...thermal head correctly use a cloth slightly dampened with alcohol or an equivalent solution with the device off Proceed as follows open the paper feed pass the cloth over the dots of the thermal head...

Страница 23: ...R 1200 VIEW 69701441 DISPLAY GLASS FOR AR 1200 VIEW 69701492 ELECTRONIC KEYBOARD BOARD AR 1200 VIEW EXT 69701502 MOTHER BOARD IRDA AR1200VIEW EXT C LINE 69701502E MOTHER BOARD IRDA AR1200VIEW EXT C LI...

Страница 24: ...s the insulation required for connection with medical systems The communication with a PC occurs through a proprietary protocol a standard IRDA protocol cannot be used et medical devices has validated...

Страница 25: ...ime constant 3 3 s Acquisition 11 bit 1000 samples s channel printing and filters 500 samples s channel in calculation and filters Resolution 5 V bit Leads 9 9 9 0 9 8 acquired 4 derived III aVR aVL a...

Страница 26: ...g in real time Automatic simultaneous acquisition ECG Autotimer acquisition at user defined intervals Paper Saving signal acquisition in simultaneous mode with possibility of copying or storing on PC...

Страница 27: ...storage Environment temperature from 10 C to 40 C Relative humidity from 10 to 95 without condensation Atmospheric pressure from 500 to 1060 hPa Dimensions 320 x 72 x 240 mm length x height x depth W...

Страница 28: ...onductive material and earthed The person performing the repair must be earthed with a suitable protective bracelet compliant with safety regulations The material must be contained in suitable antista...

Страница 29: ...29 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 16 APPENDIX B 16 1 Figures and illustrated tables 0 D...

Страница 30: ...DE VIEW VISTA LATERALE SIDE VIEW INGRESSO PAZIENTE PATIENT INPUT TASTO RESET RESET KEY 1 1 2 3 4 BATTE RY PACK 1 2 VITI FISSAGGIO FONDO MOBILE LOWER HOUSING FIXING SCREWS VITE FISSAGGIO SPORTELLO VANO...

Страница 31: ...ORE TESTA TERMICA THERMAL HEAD FLAT CONNECTOR FLAT CONNETTORE DISPLAY DISPLAY FLAT CONNECTOR 6 8 7 9 FLAT CONNETTORE TASTIERA KEYBOARD FLAT CONNECTOR INTERFACCIA SERIALE I R SERIAL INTERFACE I R FLAT...

Страница 32: ...I FISSAGGIO SCHEDA MADRE MOTHER BOARD FIXING SCREWS 2 1 3 4 5 VITI FISSAGGIO SCHEDA TASTIERA KEYBOARD FIXING SCREWS SCHEDA TAST IERA KEYBOARD VITI FISSAGGIO SISTEMA SCRIVENTE WRITING SYSTEM FIXING SCR...

Страница 33: ...COLARI DETAILS VIEW CHIMOGRAFO MOTOR ASSEMBLY T3 FLAT STAMPANTE FLAT THERMAL HEAD SPORTELLO CARTA PAPER FEED SISTEMA SCRIVENTE WRITING SYSTEM RULLO GUIDA CARTA PAPER GUIDE ROLLER MOZZO PORTA CARTA PAP...

Страница 34: ...a testa termica ruotando in senso orario o antiorario l apposito ingranaggio max 1 giro dell attrezzo Nel solo caso di stampa non uniforme procedere come segue WRITER MOUNTING INSTRUCTIONS Note Remove...

Страница 35: ...N 22 e 23 di J49 STROBE 1 e 2 STROBE 3 e 4 PIN 7 DI J49 LATCH tsetup LATCH thold LATCH thold DATA tsetup DATA 333 nSEC DATA PIN 6 DI J49 CLOCK 3MHz 1 msec TEST POINT TESTA TERMICA THERMAL HEAD TEST PO...

Страница 36: ...DIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 17 Revision Sheet Revision Sheet Service Manual ECG AR 1200 VIEW Lang English CODE 80509531 80509532 REV OM DESCRIPTION DATE...

Страница 37: ...an et medical devices SpA brand w w w c a r d i o l i n e b i z DIAGRAMS LIST ECG ar 1200 view CARDIOLINE 115 V 115 V 50 60 Hz Code 80509531 80509532 3 0 1 0 0 1 0 2 222 77777777777777 222 222 222 222...

Страница 38: ...38 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z...

Отзывы: